39
38
P6.0: Resetting to factory default settings
Available in pH 310 meter only
In the pH 310 meter, Program 6 lets you reset all parameters to factory default
settings. This clears all calibration data, memory, and any other setup functions you
might have changed.
MEAS
SETUP
A
MEAS
SETUP
B
8.8
From measurement mode
1.
Press Setup key to enter Set Up mode.
2.
Press the
▲
and
▼
keys to scroll through
subgroups until you view parameter
P6.0.
See figure
3.
Press the ENTER key to enter
parameter P6.0.
See figure
4.
Press the
▲
and
▼
keys to toggle
between NO and YES.
• NO retains current settings
• YES resets to factory default settings
5.
Press the ENTER key to confirm selec-
tion and to return to the subgroup menu.
Press the CAL/MEAS key to return to
measurement mode.
A
B
11.
Press the ENTER key to confirm the
date and move to “hour” selection.
The “hour” digits will flash.
See figure
12.
Press the
▲
and
▼
keys to toggle to
the correct hour. Note the “AM” and
“PM” indicator on the lower portion
of the display.
13.
Press the ENTER key to confirm the
hour and move to “minute” selection.
The “minute” digits will flash.
See figure
14.
Press the
▲
and
▼
keys to toggle to
the correct minutes.
15.
Press the ENTER key to confirm the
minutes and move to “second” selec-
tion. The “second” digits will flash.
See figure
16.
Press the
▲
and
▼
keys to toggle to
the correct seconds.
17.
Press the ENTER key to confirm
the seconds and return to “century”
selection.
18.
Press the CAL/MEAS key to return
to the subgroup menu. Press the
CAL/MEAS key again to return to
measurement mode.
Notes
Press the CAL/MEAS key at any point
while setting the time to return to the
subgroup menu.
SETUP
AM
F
F
SETUP
AM
G
G
SETUP
AM
H
H