To take readings:

1.

Rinse the probe with deionized or
distilled water before use 
to remove any
impurities adhering to the probe body.
If the pH electrode has dehydrated, soak
it for 30 minutes in OAKTON electrode
storage solution or a 2M–4M KCl
solution.

2.

Press ON to switch on meter. The
MEAS annunciator appears on the top
center of the LCD. The ATC indicator
appears in the lower right hand corner
to indicate Automatic Temperature
Compensation (See page 19 to set
Manual Temperature Compensation).

See figure 

3.

Dip the probe into the sample.

When dipping the probe into the sample,
the sensor or the glass bulb of the electrode
must be completely immersed into the
sample. Stir the probe gently in the sample
to create a homogenous sample.

4.

Allow time for the reading to stabilize.
Note the reading on the display.

5.

To toggle between pH and mV (or Rel
mV) readings, press the MODE key.

21

20

Taking Measurements

Be sure to remove the electrode soaker bottle or protective rubber cap on the 
electrode before measurement.

pH

ATC

MEAS

C

A

A

5.3

Taking measurements with READY indicator selected on

If the READY indicator has been activated, the READY annunciator lights when 
the reading is stable*. Switch the READY indicator on or off in Set up program
P4.1—see page 32 for directions.

*The READY indicator appears and the reading holds until the measured value exceeds the tolerance 

(± 0.02 pH; ± 0.8 mV <400; ±1.2 mV >400). Then, the READY annunciator turns off.

Taking measurements with the auto endpoint feature selected on

NOTE: this feature is available on model pH 310 only.

When a reading is stable for more than 5 seconds, the auto endpoint feature will
automatically “hold” the reading. The “hold” indicator appears on the left side of
the display. Press the HOLD key to release the reading. Switch the Auto endpoint
feature on or off in Set up program P4.1—see page 32 for directions.

http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products

Tel: (281) 359-8538  Toll Free:(866) 433-6682

Содержание pH 310

Страница 1: ...eter pH 300 Series WATERPROOF MODE HOLD ENTER ON OFF CAL MEAS MI MR SETUP pH ATC READY MEAS C 00806 23 68X248903 IP 67 Notes Write down the name and information of your OAKTON distributor here 99 Wash...

Страница 2: ...ed setup mode overview 26 27 pH 300 meter 26 pH 310 meter 27 8 2 Program 1 0 Memory clear 28 8 3 Program 2 0 Viewing previous calibration data 29 8 4 Program 3 0 Viewing electrode data 30 31 P3 1 Elec...

Страница 3: ...ctions pH ATC READY MEAS CAL pH HOLD ERR F R mV SETUP MEM ON OFF MEM AM PM C 2 1 Primary display 5 6 Secondary display 3 2 18 17 11 15 14 13 9 1 16 4 12 8 7 pH mV C Data Meter pH 300 Series WATERPROOF...

Страница 4: ...de Press CAL MEAS to return to main menu from sub menus Press CAL MEAS again to return to measurement mode from main menu ENTER Press to confirm values in Calibration mode and to confirm selections in...

Страница 5: ...t on the meter Make sure the slots of the connector are in line with the posts of the socket Rotate and push the connector clockwise until it locks See figure below 2 To remove probe push and rotate t...

Страница 6: ...d holder into the slot of the first holder until the tops of the holders are aligned and secure Inserting the Electrode into the Electrode Holder Do not use excessive force when inserting electrodes i...

Страница 7: ...t the meter off in mV units the meter will read mV units when you switch the meter on Be sure to remove the protective electrode storage bottle or rubber cap of the probe before calibration or measure...

Страница 8: ...nse solution and place it in the next pH buffer 8 Follow steps 5 to 8 for additional calibration points 9 When calibration is complete press CAL MEAS to return to pH measurement mode Notes To exit fro...

Страница 9: ...A A Relative mV Calibration 1 While in the measurement function press MODE to enter the mV mode The mV indicator appears in the upper right hand corner See figure 2 Press the CAL MEAS key The CAL ind...

Страница 10: ...al temperature compensation value DO NOT press ENTER in step 6 Press CAL MEAS instead Manual Temperature Compensation CAL C A A 5 2 5 Measurement This meter is capable of taking measurements with auto...

Страница 11: ...aking Measurements Be sure to remove the electrode soaker bottle or protective rubber cap on the electrode before measurement pH ATC MEAS C A A 5 3 Taking measurements with READY indicator selected on...

Страница 12: ...he reading To switch on or off the auto endpoint feature see Set up program P4 1 on page 32 23 22 7 Memory functions Memory Input Your meter stores data in sets pH and temperature mV or relative mV an...

Страница 13: ...mory location screen pH 310 meter only Press the ENTER key again to view the date and time the reading was taken Press the ENTER key again to return to the memory location screen See figure 4 The disp...

Страница 14: ...pH electrode offset P3 2 pH electrode slope In mV or relative mV measurement mode P3 1 relative mV offset P1 0 Memory clear P1 0 Clear all stored readings P2 0 Viewing previous calibration data P2 1...

Страница 15: ...e a new series of values This lets you avoid confusing the old values with the new ones NO is the default setting NOTE Selecting YES will wipe out all memory 8 2 SETUP MEM A A SETUP MEM B B From measu...

Страница 16: ...30 P3 0 Viewing electrode data Program 3 has two view only options that lets you check the electrode parameters for diagnostic purposes pH SETUP A A pH mV SETUP B B pH SETUP D D pH SETUP E E 8 4 P3 2...

Страница 17: ...off 33 32 P4 0 Unit configuration P4 1 READY indicator and auto endpoint function Program P4 1 lets you select READY indicator on to indicate when the reading is stable or select READY indicator off...

Страница 18: ...06 4 65 6 79 9 23 12 75 5 Press the ENTER key to confirm selection and to return to the subgroup menu Press the CAL MEAS key to return to measurement mode P4 4 Selecting C or F Available in pH 310 met...

Страница 19: ...k This helps you meet GLP Good Laboratory Practice standards 8 7 SETUP A SETUP B From measurement mode 1 Press Setup key to enter Set Up mode 2 Press the and keys to scroll through subgroups until you...

Страница 20: ...confirm the date and move to hour selection The hour digits will flash See figure 12 Press the and keys to toggle to the correct hour Note the AM and PM indicator on the lower portion of the display 1...

Страница 21: ...ion Also you can store in a pH 4 buffer with 1 100 part of saturated KCl Other pH buffers are OK for storage but NEVER use distilled water for storage After measuring 1 Rinse the pH electrode and refe...

Страница 22: ...If error persists or the meter shows incorrect values return the meter For a complete diagram of the display see page 5 ERR 1 in primary display A A 10 Troubleshooting Problem Cause Solution Power on...

Страница 23: ...ATC probe or when using with an OAKTON All in One combined pH and temperature probe See section 4 5 Temperature calibration page 17 Continued on next page 12 Specifications Mode pH Temperature mV 10...

Страница 24: ...areless or insufficient packing A restocking charge will be made on all unauthorized returns NOTE We reserve the right to make improvements in design construction and appearance of products without no...

Отзывы: