Instruction Manual
PD 300
8.5
P4.0: Resetting to factory default settings (pH)
This program lets you reset all pH parameters to factory default settings. This clears all
calibration data and any other pH setup functions you might have changed.
The following settings will remain as you have set them:
•
Temperature unit of measure (°C or °F)
•
The temperature offset calibration value
•
All DO calibration data and parameters
There are also some other parameters that retain settings when reset is done. See Addendum
3 on page 68 for a table of factory default settings.
From measurement mode
Figure 27 – P4.0: Reset to factory
default values
1. Press
the
MODE
key to select pH measurement mode.
2.
Press SETUP key to enter Set Up mode.
3. Press
the
▲
or
▼
keys to scroll through subgroups until
you view parameter P4.0 in the lower display.
4.
Press the ENTER key to enter parameter P4.0.
5. Press
the
▲
or
▼
keys to toggle between NO and YES.
See figure on right.
•
NO retains current settings
•
YES resets to factory default settings.
6. Press the ENTER key to confirm selection and to return
to the measurement mode. Otherwise press CAL/MEAS
key to return to measurement mode without resetting to
factory default.
NOTE
See page 68 for a table of factory default settings.
36