Instruction Manual
PD 300
7. Rinse the electrode with deionised water or in
rinse solution, and place it in the next pH buffer.
8. Follow steps 5 to 8 for additional calibration
points. See Figure 7.
9. When calibration is complete, press
CAL/MEAS to return to pH measurement mode.
NOTE
To exit from pH calibration mode without confirming
calibration, DO NOT press ENTER in step 6. Press
CAL/MEAS instead.
If the selected buffer value is not within
±
1.0 pH from the
measured pH value: the electrode and buffer icon blink
and the ERR annunciator appears in the lower left
corner of the display. See Figure 8.
Figure 7: Next point calibration for
pH 4.01
To limit the number of pH buffer values available during
calibration, see Setup Section 8.4
P3.2: Selecting
number of pH calibration points
on page 34.
Figure 8: Err message and
electrode icon will appear if
incorrect buffer is used
11