1065O101_MAN_ORP5_35634-54 March 2018
www.4oakton.com
Sensor Replacement
Screw off the sensor ring, unplug the sensor, plug in the new
replacement sensor (pay attention to the probe’s position), and
rescrew on the sensor ring
Battery Installation
The tester uses four AAA batteries. Please install batteries according to
the following steps. Note the correct direction of battery installation: the
positive side (+) of every single battery must face up. Incorrect installa-
tion of batteries will cause damage to the tester and create a potential
hazard.
1. Pull the battery cap up.
2. Slide the battery cap along the direction of arrow.
3. Open the battery cap.
4. Insert the batteries
(ALL POSITIVE SIDES FACING UP).
5. Close the battery cap.
6. Slide and lock the battery cap along the direction of arrow.
7. Fit the tester’s cap while making sure to push all the way
down. The tester’s waterproof design may be compromised if
the cap is not fitted correctly.
4 x AAA Batteries
+
+
+
+
_ _
_ _
1
2
3
4
5
6
7
Warranty
We warrant this instrument to be free from defects in material and
workmanship and agrees to repair or replace free of charge, at
option of Oakton Instruments, any malfunctioned or damaged
product attributable to responsibility of Oakton Instruments, for a
period of
two years
from the delivery (a
six-month
limited warranty
applies to sensors). This warranty does not apply to defects resulting
from actions such as misuse (violation of the instructions in this
manual or operations in the manner not specified in this manual),
improper maintenance, and unauthorized repairs. Warranty period is
the time limit to provide free service for the products purchased by
customers, not the service life of the tester or probe.
Oakton Instruments reserves the right to update the information in
this manual without giving notice in advance.
Self-Diagnostic Messages
Symbol
Self-diagnosis
information
How to fix
Er2
CAL/
is pressed
before measurement
is stable (
appears)
Wait for the measurement stability
icon (
) to appear and stay, then
press CAL/
.
Specifications
ORP (mV)
Range
±1000 mV
Resolution
1 mV
Accuracy
±0.2% full-scale
Temperature
Range
32 to 122°F (0 to 50°C)
Resolution
0.1°F/°C
Accuracy
±0.9°F (0.5°C)
Display:
LCD with three-color backlight. Blue = measurement;
Green = calibration; Red = alarm
Reading lock:
HOLD icon
Power:
four AAA batteries (included); >400 hours of continuous
operation
Low-voltage warning:
battery status icon flashes
Auto power-off:
tester automatically turns off after 8 minutes
of nonuse
IP rating:
IP67 (waterproof), floats on water when sensor cap is on
Dimensions
(L x W x H): 7" x 1.5" x 1. 5" (17.8 x 4 x 4 cm)
Weight:
4.7 oz (133 g)
Ordering Information
Model
Product description
Catalog number
ORP5
ORPTestr 5 pocket tester
35634-54
—
Replacement sensor for ORPTestr 5
35634-58