Figure 14 – Central Unit Option Warning Screen
Ensure that there is no waste steam which is came from the external steam generator.
Steam supply minimum temperature and pressure values should be 139°C and 2.6 Bar;
maximum temperature and pressure values should be 143°C and 3.0 Bar.
After the generator option is selected, the device power should be turned off-on.
In case steam value is below the determinated minimum temperature and pressure
values or above the determinated maximum temperature and pressure values,
undesired failures can be occurred on the device.