TEXTILES 

While preparing textiles for treatment in the sterilizer, care must be taken that the folds in the 
textiles  are  arranged  in  parallel,  and  that  the  items  are  packed  side  by  side.  This  vertical 
configuration ensures that channels can form between the textile folds for the air to flow out and 
steam to flow in. Do not stack textiles on the top each other as this hinders the penetration of 
steam into the packages of textiles. 

 

Figure 7 - Loading textiles properly 

When loading sterilization containers with textile items, care should be taken ensure that they 
retain their vertical orientation, but that the items are not squashed together. This would prevent 
the  formation  of  flow  channels  for  air  and  steam.  If  the  packages  of  textiles  cannot  be  kept 
upright, then it might be advisable to wrap them in sterilization paper. 

The textiles must not touch the sides or the base of the sterilization container, since they might 
become saturated with condensate. 

For good drying results, the textiles should also be as dry as possible when they are placed in the 
autoclave.  The  heat  stored  in  the  chamber  and  sterilization  container  may  not  otherwise  be 
sufficient to evaporate both the moisture and the condensation. 

6.4.3.

 

INSTRUMENTS 

Where appropriate, instruments should be disassembled before placing them in the autoclave, 
as this will improve the drying results. 

The use of lubricants (such as instrument oil) should be avoided unless absolutely necessary. Prior 
confirmation  should  be  obtained  from  the  manufacturer  of  such  agents  that  they  are  in  fact 
suitable for steam sterilization. Substances which are hydrophobic or impenetrable for steam can 
not only lead to poor drying results, but may also mean that the steam sterilization is unsuccessful 
, since not only the instruments are protected but also micro-organisms. 

6.4.4.

 

LOADING THE AUTOCLAVE 

Textiles and instruments should not be sterilized together in one sterilization container. Textiles 
and instruments in separate sterilization containers should as far as possible not be sterilized in 

Содержание NC 300

Страница 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A NC 300 NC 430 NC 570 NC 710 STEAM STERILIZERS USER S MANUAL 1984 Z14 K25 285 Rev No 08 Rev Date 01 2018...

Страница 2: ...perate damaged instrument 4 do not operate instrument with a damaged cable 5 do not move instrument during operation In case of a problem contact your N ve agent for an authorized service or maintenan...

Страница 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Страница 4: ...TO WATER SUPPLY 15 5 7 3 CONNECTION TO AIR SUPPLY 15 6 OPERATING PRINCIPLES 15 6 1 PRIOR TO OPERATION 15 6 2 OPERATION PHASES 23 6 3 SAMPLE REPORT 24 6 4 INSTRUCTIONS ON DRYING 25 6 4 1 DRYING IN STER...

Страница 5: ...situated outside the useful volume of the unit maintains a homogenous temperature distribution within the sterilization chamber This jacket is also used to pre heat system prior to sterilization thus...

Страница 6: ...ure 144 C Temperature Sensors PT 100 Test Programs Vacuum Bowie Dick Max chamber pressure 3 0 bars Max Jacket Pressure 3 0 bars Max Generator Pressure 3 0 bars Volume Liter 300 430 570 710 STU capacit...

Страница 7: ...hould have a boiling point of 100 C at sea level at 760 mmHg 1 atm atmospheric pressure Proper sterilization program shall be selected in accordance with the load type and the sterilization load shall...

Страница 8: ...ture up to 31 C 80 Maximum altitude 2000 m Temperature for maximum performance 15 C 25 C 5 2 HANDLING AND TRANSPORTATION All handling and transportation must be carried out by using proper equipment a...

Страница 9: ...sers should be chosen according to weigh of the device The weight of the device with pallet NC 300 950 kg NC 430 1050 kg NC 570 1200 kg NC 710 1500 kg Unfix 4 fixing screws which are under the pallet...

Страница 10: ...ric air water and drain connections Figure 2 5 3 UNPACKING Remove the cardboard box packing and the second nylon wrapping around the instrument Ensure that no damage has occurred during transportation...

Страница 11: ...circuit breaker should be used for protection against indirect contact in case of isolation fault 5 5 POSITIONING Check that the positioning is suitable for the users Check that the instrument is stab...

Страница 12: ...12 5 6 GENERAL PRESENTATION Figure 3 1 2 3 4 5 6 7 8...

Страница 13: ...uttons on top are called touchscreens These buttons are activated by patting the requested commands shown on the screen It is the interface utilized by the user to communicate with the control unit th...

Страница 14: ...e process is finalized by pressing the button once more Air Intake Button When the chamber is under vacuum a negative value is displayed by the chamber manometer The chamber vacuum is broken by taking...

Страница 15: ...shall be connected to the water inlet of the unit after being treated by a suitable water softening unit It shall be always kept in mind that the supply water quality has a direct effect on the lifet...

Страница 16: ...16 Password screen is accessed when Start is pushed When the password is entered correctly program screen appears The password for first usage is 1111...

Страница 17: ...17 When TEST PROGRAMS is pushed the page where two test programs are listed is shown When PRESET PROGRAMS is pushed the page where preset programs are listed is shown...

Страница 18: ...18...

Страница 19: ...2 15 3 5 3 Vacuum test The corresponding program screen is displayed when a program is selected On this page the program line displays the selected program and the total time dissipated after program...

Страница 20: ...to function the unit when it is in stand by or to stop it while it is functioning The Settings item which is the third choice on the Menu page takes the user to the Settings screen The user can adjust...

Страница 21: ...In order to run a private program select PRIVATE PROGRAMS in programs menu page Liquid program is a private program therefore success of sterilization and damages that occurs on the materials which de...

Страница 22: ...ar is pushed on menu page the page where the temperature and pressure values are listed appears When CALIBRATION is pushed on menu page the page to do calibration for pressure and temperature values i...

Страница 23: ...nces for the built in programs with a pre vacuum phase The pressure in the chamber is lowered below ambient pressure by removing the air out of the chamber and steam is injected in to replace the volu...

Страница 24: ...a steam sterilization system 6 3 SAMPLE REPORT NUVE STEAM ART NC 430 SERIAL NUMBER 03 0025 F RM NUVE A S ADDRESS ANKARA TELEPHONE 312 3992830 DATE 30 10 2013 TIME 12 54 51 Program WRAPPED POROUS MATE...

Страница 25: ...s these This leads to steam condensing on the instruments and containers Figure 4 Some of the condensation drops to the bottom of the sterilization container Figure 5 Figure 4 Formation of condensatio...

Страница 26: ...might become saturated with condensate For good drying results the textiles should also be as dry as possible when they are placed in the autoclave The heat stored in the chamber and sterilization co...

Страница 27: ...t be at the bottom Figure 8 Loading the autoclave 6 4 5 LOADING CONTAINERS WITH SOFT STERILIZATION PACKAGES Soft sterilization packages such as paper bags or transparent sterilization packages can be...

Страница 28: ...nce in five years by the authorized personnel After each 2000 runs the instrument should be controlled by authorized technical service personnel The door gasket shall be replaced by the authorized per...

Страница 29: ...or regulations by an authorized inspector According to calculations number of allowable cycles for the operation conditions are 10 000 at pressure fluctuation between 0 bar to 2 05 bar and 20 000 at...

Страница 30: ...The mains supply is present Keep in mind that the steam in the sterilization chamber is not discharged upon power failure If steam is to be discharged open the safety valve on top of the unit to relea...

Страница 31: ...f the temperature and pressure sensors which perform the test functions of the equipment are defected Error 06 Heater Failure The error code is displayed when the thermic relays of the heaters give a...

Страница 32: ...imit The pressure has not reached the predetermined value in the predetermined time during air in phase Error 15 Steam Supply to Gasket Time Exceeded Gasket pressure has not reached the predetermined...

Страница 33: ...followings changes Connect the device and the steam line which is came from the external steam generator Choose Steam Generator Central option by following Settings Maintenance pages Figure 13 Mainten...

Страница 34: ...ues should be 139 C and 2 6 Bar maximum temperature and pressure values should be 143 C and 3 0 Bar After the generator option is selected the device power should be turned off on In case steam value...

Страница 35: ...35 11 ELECTRICAL CIRCUIT DIAGRAM...

Отзывы: