8.
DISPOSAL MANAGEMENT CONCEPT
The currently valid local regulations governing disposal must be observed. It is in the responsibility
of the user to arrange proper disposal of the individual components.
All parts which may comprise potentially infectious materials have to be disinfected by suitable
validated procedures (autoclaving, chemical treatment) prior to disposal. Applicable local
regulations for disposal have to be carefully observed.
The instruments and electronic accessories (without batteries, power packs etc.) must be
disposed off according to the regulations for the disposal of electronic components.
Batteries, power packs and similar power source have to be dismounted from electric/electronic
parts and disposed off in accordance with applicable local regulations.
9.
TROUBLESHOOTING
If the device fails to operate, please check the followings:
The power switch is on;
the automatic fuse on the unit board is on and not defective;
The plug is plugged-in properly;
The plug is not defective;
The mains supply is present.
Keep in mind that the steam in the sterilization chamber is not discharged upon
power failure. If steam is to be discharged, open the safety valve on top of the unit to
release the steam inside.
9.1.
WARNING MESSAGES AND EXPLANATIONS
Door Open: This warning is displayed if the user presses “START” before closing the door. If the
message is displayed when the door is closed, then there is a failure in the system.
Insufficient Water, Water Filling Failure: This warning is displayed when there is a shortage in
the water mains or the water in the storage tank is insufficient. If the message is displayed despite
there is no such problem; then there is a failure in the system.
Change Filter: A long air intake duration beyond acceptable limits at the end of the drying period
means that the filter is blocked or the expiry date is exceeded. Change the filter. If this message
is displayed with a new filter installed; then there is a failure in the system.
9.2.
ERROR CODES
The Error Codes may appear before a program start, after the sterilizer is powered on.
The Error Codes may appear immediately after a program started or during a run.
These messages are accompanied by an alarm tone.