23
Ensure that power led is turned on (See Figure 8). Power led is on when energy is supplied to
the GSM module. SMS led starts to flash while the module sending SMS.
Connect the antenna cable to antenna inlet on the GSM module (See Figure 7).
Antenna has magnet to place it easily. Place the antenna on a place where the signal of GSM
module is high.
If the GSM module is not connected or does not send messages although it is
connected, “modem” error code appears in the error history. If the GSM module is
connected and cannot send messages, “SMS” error code appears in the error
history. Modem and SMS errors do not appear when SMS submenu on the SMS page
is selected as “off”.
10.2.
NüveWarn
™
REMOTE ALARM SYSTEM
KN series blood bank refrigerators have “remote alarm” feature. This feature can be used only if
an optional “remote alarm kit” (K 13 009) is connected to the refrigerators. The remote alarm
will be activated for all alarm situations as power failure, open door, broken temperature
sensor,
2°C deviation of chamber temperature from set value.
10.2.1.
MOUNTING REMOTE ALARM
The remote alarm kit includes an alarm unit and 10 meters of cable with a connector. Connect
one side of the cable to the connector at the rear of the refrigerator and the other side to the
alarm unit.
Mount the remote alarm kit to a wall or board not more than 10 meters away from
the refrigerator where it is easily accessible and observable.
Figure 8