13
This symbol denotes enter key. It is used for approval of adjustments.
The key corresponding to this symbol is used to turn the illumination on or off. If
the key is green, it shows that the lamp is off. The lamp goes on when the key is
pushed and the key turns red. Lamp goes off when the red key is pushed.
5.8.
PRIOR TO OPERATION
Plug-in the instrument to correctly grounded sockets.
Insert the power switch key to the on/off lock on the control panel.
Turn the device on by rotating the key clockwise direction.
See that the microprocessor control system activates.
Learn the functions of the control panel (See Section 5.7).
6.
OPERATING PRINCIPLES
6.1.
PROGRAMMING
For temperature homogeneity, never take out the plexiglas dividers inside the
drawers and do not overload the drawers.
6.1.1.
OPERATING MODE
When the refrigerator is powered on, it starts to
operate automatically according to the last set
values. The screen on the left is operating mode
screen.
Set
temperature
and
chamber
temperature values appear in this screen. Current
time and date are shown at the top line. “
”
symbol flashes on this screen while refrigerator
performs cooling.
Password screen will appear when the MENU
button (F1) is pushed. You can enter the password
by using the value increase (F1) and decrease keys
(F2). The password is user’s password and can be
changed on settings page (See section 6.1.2.). The
password is 0000 for the first usage.