Fill Oxygen Analyzer - Alternate Calibration Method
The Fill Oxygen Analyzer can also be calibrated in ambient air as an
alternative. This is especially useful during routine re-calibration
while the system is operating. Different settings may be used
depending on location, so verify your actual ambient conditions and
refer to the Oxygen Analyzer manual for details.
1) Remove the Flow Adapter Cap covering the Sensor.
2) Expose the Sensor to ambient air for approximately 15 seconds.
3) Adjust Calibration Knob until Display reading stabilizes at 20.9%.
4) Reinstall the Flow Adapter Cap to the analyzer.
5) The Fill Oxygen Analyzer is now ready for use.
14.4
Attaching Scuba Cylinder
Only one fill whip is attached to a Scuba Cylinder at this point. The second whip will continue
to be used to control HP Compressor outlet pressure.
Attach one HP Compressor Fill Whip to a Scuba Cylinder.
Leave Cylinder Valve closed.
Additional cylinders or storage tanks can be attached after
the system is ready to pump Nitrox.
High pressure cylinders that are filled quickly will become hot and due to the increased
internal temperature the cylinder pressure will increase. This will leave a diver with less
pressure inside the cylinder once cooling has occurred. This will decrease the amount of time
the diver may spend underwater which may be critical during a deep dive. Customers must be
warned of this possibility if cylinders are delivered for use while warm. Always fill all breathing
gas cylinders slowly to avoid overheating.
Sensor Cap
Removed
Calibration
Knob
Fill Oxygen Analyzer
Содержание Open Voyager II
Страница 1: ...User Manual Voyager II Open Voyager II Nitrox System Rev 03 19...
Страница 17: ...Voyager II Open Voyager II Nitrox System Nuvair Page 17 www nuvair com...
Страница 46: ...Voyager II Open Voyager II Nitrox System Nuvair Page 46 www nuvair com NOTES...
Страница 47: ...Voyager II Open Voyager II Nitrox System Nuvair Page 47 www nuvair com...