18
10. Specifications
10.1
Power Requirements
Voltage ±15Vdc
-
±18Vdc
Current
±5A rms, 10A peak
NOTE: A power supply with sufficient output current should be used. The output current
required depends on the supply voltage, galvanometer model, mirror load, and command
waveform and duty cycle. Lower supply voltages, larger galvanometers, larger mirrors,
and more demanding waveforms and duty cycles will require more current. It is a good
practice to setup your application with a laboratory power supply and measure the
required current during the most demanding application your system can run. Then
select a power supply appropriate for your application.
Select a high quality power supply with good clean DC outputs. Power supplies with
excessive ripple or noise will cause scanning problems in your system.
The electronics internal to the 3XB have generous output capacitors that supply
instantaneous peak current to the output drive amplifiers, and lower the peak output
requirements of your system power supply.
10.2
Environmental
Operating Temperature:
0
°
C to +45
°
C (with appropriate cooling)
Storage Temperature:
-10
°
C to +60
°
C
Humidity:
Non-Condensing
Temperature Stability:
10 PPM per
°
C Standard
10.3
Size and Weight
Size
484.58mm (19.05in) x 246.87 (9.68in) x 215.7 (8.49in)
Overall - See Outline drawing for specific information
Weight 13.6kg (30.0 lbs)
Содержание 3XB
Страница 1: ...Nutfield Technology Inc 1 Wall St Suite 113 Hudson NH 03087 USA 91 0017 Rev E 3XB 3 Axis User Manual...
Страница 2: ...2 This Page Intentionally Left Blank...
Страница 14: ...14 f Click the Advanced Button Under the Optic Tab...
Страница 19: ...19 11 0 Outline Drawing...
Страница 22: ...22 Appendix A CorCalcII Information A 1 10mm Aperture 266nm NOTE 6 0mm Entrance Beam Required...
Страница 23: ...23 A 2 10mm Aperture 355nm NOTE 1 31mm Entrance Beam Required...
Страница 24: ...24 A 3 15mm Aperture 532nm NOTE 4 0mm Entrance Beam Required...
Страница 25: ...25 A 4 30mm Mirror Aperture 1064nm NOTE 8 10mm Entrance Beam Required...
Страница 26: ...26 A 5 30mm Aperture 10 6um NOTE 11 55mm Entrance Beam Required...
Страница 27: ...27 A 6 45mm Aperture 10 6um NOTE 14mm Entrance Beam Required...
Страница 28: ...28 A 7 30mm Aperture Dual Wavelength 1 of 2 532nm NOTE 6 1mm Entrance Beam Required...
Страница 29: ...29 30mm Aperture Dual Wavelength 2 of 2 1064nm NOTE 6 1mm Entrance Beam Required...
Страница 30: ...30 A 8 30mm Aperture 355nm NOTE 3 5mm Entrance Beam Required...
Страница 31: ...31 A 9 30mm Aperture 532nm NOTE 6 0mm Entrance Beam Required...