8.2.2.3 Option Setup
The option set up menu allows
A QUALIFIED TECHNICIAN
to configure several different optional parameters per
the menu below. Each parameter sub-menu will be described as well as the display will the default conditions as
shown
•
Alarm Setup
•
Alarm Silence:
This parameter allows for the selection of the alarm silence key function. When the
alarm silence function is on, all current and future alarms will be silenced for the designated alarm
silence time (i.e. default time is 15 minutes). When the alarm silence function is off, all current alarms
will be silenced for the designated alarm silence time. If a new alarm is present, the audible alarm will
again be turned on.
Содержание LabGard NU-560-400
Страница 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Страница 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Страница 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Страница 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Страница 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Страница 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Страница 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Страница 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Страница 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Страница 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Страница 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Страница 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Страница 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Страница 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Страница 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...
Страница 102: ...OM0295 Rev 2 September 2020 Page 102 of 103...
Страница 103: ...OM0295 Rev 2 September 2020 Page 103 of 103...