OM0295
Rev 2 September 2020
Page 71 of 103
8.1 Error Message Troubleshooting Guide
Error Message
Indicator
Correction
Window Alarm
(Window High)
Sliding window is above its standard working height
or micro switch is not operating properly.
Verify standard working height and window micro
switch operation.
Airflow Alarm
Red Downflow Arrow
(Downflow Low Limit)
Downflow airflow fell below its lower limit alarm
setpoint.
Re-certify cabinet to proper airflow setpoints.
Airflow Alarm
Red Downflow Arrow
(Downflow High Limit)
Downflow airflow went above its high alarm
setpoint.
Re-certify cabinet to proper airflow setpoints.
Airflow Alarm
Red Inflow Arrow
(Inflow Low Limit)
Inflow airflow fell below its lower limit alarm
setpoint.
Check orientation of exhaust sensor shroud.
Re-certify cabinet to proper airflow setpoints.
Airflow Alarm
Red Inflow Arrow
(Inflow High Limit)
Inflow Airflow went above its high alarm setpoint.
Check orientation of exhaust sensor shroud.
Re-certify cabinet to proper airflow setpoints.
Low Pressure Alarm
(Low pressure Limit)
Indicates low pressure or low cabinet airflow
Re-certify cabinet to proper airflow setpoints.
Cabinet fluorescent
lights won't turn on.
Check light fuse on main control board.
Check fluorescent lamps.
Check voltage coming out of main control board to
light ballasts.
Check light starters, if present.
Check ballast.
Cabinet Blower
won't Turn On.
Check blower fuse on main control board.
Check voltage coming out of main control board.
Check wiring to blower. Check blower motor. (Note:
blower motor has internal thermal protector. Let
blower motor cool off for a minimum of 30 minutes
to assure thermal protector is not open).
Display indicates
(Remote Override
Active)
Indicates that the remote override is activated,
preventing the usage of the cabinet.
Power Loss Alert!
Indicates a power interruption has occurred.
Press DISPLAY to clear message.
Cabinet outlets won't
turn on.
Check outlet fuse located on main control board.
Check voltage coming out of main control board.
Cabinet ultraviolet light
won't turn on.
Check sliding window position so that it's fully closed.
Check blower/lights fuse on main control board.
Check voltage coming out of the main control board
to ultraviolet light ballast.
Check light starters, if present. Check ballast.
Blower or light fuse
continues to blow after
replacement.
Check for short on output of fuse.
Isolate output of fuse by disconnecting light circuit,
blower circuit, etc. to isolate short.
Replace UV Light!
Indicates that the UV light needs replacement
Replace UV light and clear UV run time clock.
Display indicates
(Night Setback Active)
Indicates that the night setback is activated,
preventing the usage of the cabinet.
Certification due in X
weeks.
Indicates the need to schedule the cabinet
certification based on the scheduled timetable
programmed.
Message will clear once the certifier updates the
current certification date.
Active Alarms
DN Sensor Comm!
EX Sensor Comm!
Indicates a digital communications error from the
main control board to the airflow sensors.
Check connectors and wires from main control board
to the airflow sensors. DN indicated downflow
sensor. EX indicates exhaust sensor.
Active Alarms
DN Sensor Error!
EX Sensor Error!
Indicates an error signal generated by the sensor.
Check airflow probe connector on main board. (Ref.
Section 7.11). Replace airflow sensor if required.
Touch Link Display User
Interface
Touch area’s of display do not Align w/ icons
Calibrate display in adjust display brightness screen
Run Time Failure
Indicates a temporary microprocessor issue
Clean alarm a see it reoccurs, check line voltage to
control board, replace control board
Содержание LabGard NU-560-400
Страница 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Страница 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Страница 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Страница 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Страница 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Страница 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Страница 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Страница 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Страница 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Страница 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Страница 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Страница 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Страница 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Страница 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Страница 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...
Страница 102: ...OM0295 Rev 2 September 2020 Page 102 of 103...
Страница 103: ...OM0295 Rev 2 September 2020 Page 103 of 103...