OM0295
Rev 2 September 2020
Page 23 of 103
6.1.4 Standby/Run Mode Alarms
If present, standby/run mode alarms will be both visual and audible, the Red LED oval under the TOUCHLINK LCD
display will turn on, and the TOUCHLINK LCD screen will also display a description of the alarm in place of the
NuAire Logo along with alarm silence icon. Depending upon the alarm type, the LFBSC profile will also indicate in
red the alarm present. Audible alarms can be silenced or will produce an alarm tone for 10 seconds, then into a
ring back cycle of once every 2 seconds. Pressing the alarm silence icon will silence the audible alarm for 15
minutes, then into a ring back cycle at of once every 2 seconds.
Alarm Types
The middle of the display just below the icons is the message area that will indicate alarms, errors or other notable
conditions. Since the message area is limited to one line of text, only the highest priority message will be
indicated. However, for alarm conditions, also displayed is a caution symbol and pressing it will display all present
alarm conditions. The list below represents the highest to lowest priority.
•
Runtime Failure
•
Power Loss Alert
•
Downflow Sensor Error
•
Inflow Sensor Error
•
Exhaust Sensor Error
•
Downflow Sensor Communication
•
Inflow Sensor Communication
•
Exhaust Sensor Communication
•
Downflow Low Limit
•
Downflow High Limit
•
Inflow Low Limit
•
Inflow High Limit
•
Exhaust Low Limit
•
Exhaust High Limit
•
Cabinet Pressure Low
•
Window High
•
Window Low
•
Replace Filter
•
Replace UV Lamp
•
Recertification Past Due
•
Recertification Due In (x) Week(s)
6.1.5 Menu Icon
The menu icon, when pressed will provide a list of menu items for various TouchLink functions.
Содержание LabGard NU-560-400
Страница 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Страница 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Страница 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Страница 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Страница 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Страница 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Страница 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Страница 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Страница 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Страница 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Страница 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Страница 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Страница 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Страница 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Страница 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...
Страница 102: ...OM0295 Rev 2 September 2020 Page 102 of 103...
Страница 103: ...OM0295 Rev 2 September 2020 Page 103 of 103...