OM0228
Page
87
of
99
February/2014
Read
Inputs
Read
inputs
all
a
service
technician
to
exercise
or
check
these
input
functions.
The
inputs
may
be
checked
by
altering
the
state
of
the
input
function
(i.e.
sliding
window
position)
and
monitoring
the
change
on
the
display.
8.3
Airflow
Sensor
Performance
Verification
The
individual
airflow
sensors
can
be
routinely
checked
during
calibration
or
in
diagnostics
to
assure
they
are
reading
and
active
within
the
range
of
use
(0
to
200
fpm)
(0
to
1.02
mps).
The
airflow
sensors
can
also
be
checked
in
the
run
mode
through
performance
verification,
for
responsiveness
to
changing
airflow
conditions.
8.3.1
Run
Mode
To
check
the
airflow
sensor
in
run
mode,
first
allow
the
cabinet
to
operate
normally
for
a
minimum
of
5
minutes.
Then,
place
a
rolled
piece
of
paper
over
the
downflow
sensor
in
the
workzone
and
leave
the
paper
on
the
sensor
for
at
least
2
minutes
and
then
remove.
This
action
will
cause
the
cabinet
to
go
into
a
downflow
alarm
condition.
The
exhaust
airflow
reading
should
increase
during
this
test.
However,
the
downflow
reading
should
go
down
below
25
fpm
(.13
m/s).
There
should
also
be
a
noticeable
increase
in
motor/blower
noise.
It
would
also
be
recommended
to
monitor
motor/blower
voltage
during
the
test.
The
motor/blower
voltage
should
be
monitored
from
when
the
cabinet
is
running
normally.
During
the
test,
when
the
downflow
sensor
is
covered,
the
motor/blower
voltage
should
be
steadily
increasing
to
slightly
under
line
voltage.
When
the
downflow
sensor
is
uncovered,
the
motor/blower
voltage
should
decrease
and
airflow
readings
should
be
within
the
calibration
range.
If
the
motor/blower
voltage
does
not
change,
an
airflow
sensor
problem
could
exist.
Please
consult
with
NuAire
Technical
Service.
MENU
xx:xx
AM
xx/xx/xx
UV
FL
KEYPAD
X = 0, Y = 0
DOWNFLOW SENSOR
60 FPM
EXHAUST SENSOR
1100 CFM
REMOTE OVERRIDE
OPEN
NIGHT SETBACK
OPEN
WINDOW POSITION
NORMAL
PRESSURE SENSOR
ACTIVE
AC LINE HZ
60 HZ
AC LINE VOLTS
115 VAC
MOTOR RPM
1000 RPM
NEXT
MENU
xx:xx
AM
xx/xx/xx
UV
FL
WINDOW HIGH LIMIT
CLEAR
WINDOW HIGH
CLEAR
WINDOW CLOSED
CLEAR
WINDOW SAFE
TRIPPED
NEXT
Содержание Labgard NU-430-400
Страница 7: ...OM0228 Page 7 of 99 February 2014 ...
Страница 9: ...OM0228 Page 9 of 99 February 2014 ...
Страница 10: ...OM0228 Page 10 of 99 February 2014 ...
Страница 16: ...OM0228 Page 16 of 99 February 2014 ...
Страница 17: ...OM0228 Page 17 of 99 February 2014 ...
Страница 41: ...OM0228 Page 41 of 99 February 2014 ...
Страница 93: ...OM0228 Page 93 of 99 February 2014 ...
Страница 94: ...OM0228 Page 94 of 99 February 2014 ...
Страница 95: ...OM0228 Page 95 of 99 February 2014 ...
Страница 96: ...OM0228 Page 96 of 99 February 2014 ...
Страница 97: ...OM0228 Page 97 of 99 February 2014 ...
Страница 98: ...OM0228 Page 98 of 99 February 2014 ...
Страница 99: ...OM0228 Page 99 of 99 February 2014 ...