OM0228
Page
62
of
99
February/2014
7.10
Main
Control
Board
Description
&
Replacement
To
access
the
main
control
board
for
fuse
or
board
replacement,
remove
screws
at
each
upper
side
of
the
control
center
and
allow
the
control
center
to
rotate
down,
resting
on
the
safety
straps.
Now
the
main
control
board
is
exposed
for
service.
7.10.1
Main
Control
Board
Replacement
The
main
control
board
consists
of
two
interconnected
Printed
Circuit
Board
(PCB)
assemblies.
The
front
PCB
contains
the
LCD
display.
The
back
PCB
contains
the
power
supply,
sensor
inputs/outputs
and
control
inputs/outputs
components.
The
mechanical
and
electrical
interconnects
for
the
two
PCB's
all
occur
within
the
assemblies
and
are
fastened
together
with
standoffs
and
screws.
7.10.2
Main
Control
Board
Fuse
Replacement
Disconnect
electrical
power
from
cabinet
before
fuse
replacement.
All
AC
circuits
are
fuse
protected
and
when
replacement
is
necessary,
USE
ONLY
FUSES
OF
SAME
TYPE
AND
RATING
FOR
PROTECTION
AGAINST
RISK
OF
FIRE
.
DESCRIPTION:
BLOWER
FUSE
OUTLET
FUSE
LIGHT
FUSE
POWER
INPUT
FUSE
FUSE
TYPE:
TIME
‐
LAG
TIME
‐
LAG
TIME
‐
LAG
SUB
‐
MIN
(250V)
FUSE
SIZE:
1/4
X
1
‐
1/4
INCH
5
X
20MM
5
X
20MM
TR
‐
5
NU
‐
430
‐
400
8
AMPS
3
AMPS
1
AMP
4
AMPS
NU
‐
430
‐
600
10
AMPS
3
AMPS
1
AMP
4
AMPS
7.10.3
Main
Control
Board
Replacement
Note:
All
setup
and
calibration
data
will
be
lost,
the
memory
reinitialized
to
the
default
values
and
all
control
functions
reset
to
an
initial
cabinet
power
condition.
If
possible,
before
the
main
control
board
replacement,
it
would
be
preferred
to
know
the
operational
parameters
of
the
cabinet,
(i.e.
motor/blower
voltage,
setpoints,
and
airflow
data
from
previous
certification.
Disconnect
electrical
power
from
the
cabinet
before
attempting
any
maintenance
action.
The
main
control
board
is
fastened
to
the
control
center
with
(6)
6
‐
32
screws.
All
electrical
connections
are
made
with
removable
terminals
and/or
Faston
connectors
except
for
the
motor/blower
connector
which
uses
a
screw
terminal.
Remove
all
electrical
connections
and
fasteners
then
remove
the
main
control
board
from
the
control
center.
Install
new
main
control
board
by
reattaching
all
electrical
connections
and
fasteners.
Once
installed,
rotate
control
center
to
normal
position
and
fasten
in
place.
Now
reconnect
power
to
cabinet.
Upon
BSCC
system
power
up,
a
system
MASTER
RESET
must
be
performed
to
clear
the
microprocessors
non
‐
volatile
memory
to
assure
proper
system
function.
CAUTION
!
Содержание Labgard NU-430-400
Страница 7: ...OM0228 Page 7 of 99 February 2014 ...
Страница 9: ...OM0228 Page 9 of 99 February 2014 ...
Страница 10: ...OM0228 Page 10 of 99 February 2014 ...
Страница 16: ...OM0228 Page 16 of 99 February 2014 ...
Страница 17: ...OM0228 Page 17 of 99 February 2014 ...
Страница 41: ...OM0228 Page 41 of 99 February 2014 ...
Страница 93: ...OM0228 Page 93 of 99 February 2014 ...
Страница 94: ...OM0228 Page 94 of 99 February 2014 ...
Страница 95: ...OM0228 Page 95 of 99 February 2014 ...
Страница 96: ...OM0228 Page 96 of 99 February 2014 ...
Страница 97: ...OM0228 Page 97 of 99 February 2014 ...
Страница 98: ...OM0228 Page 98 of 99 February 2014 ...
Страница 99: ...OM0228 Page 99 of 99 February 2014 ...