CAUTION
ō
Be sure to clean and lubricate the handpiece prior to autoclaving. If blood remains on the internal surface it can become
clotted and cause product failure.
ō
Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and
damage to the product from chemical residue on other instruments.
ō
Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
ō
To avoid product failure, do not use a sterilizer whose temperature exceeds 138°C during use including the dry cycle.
The chamber temperature of some type of sterilizers may exceed 138°C. Contact the sterilizer manufacturer for detailed
information.
ō
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free
from dust, salt and sulphur.
ō
Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile condition.
ō
Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
ō
Only Handpiece, Internal Spray Nozzle and Nozzle Holder can be autoclaved. Do not autoclave other components (Y-Connector,
Irrigation Tube, etc.).