NSK Endo-Mate TC2 Скачать руководство пользователя страница 3

English

Classification of Devices 

Classification by type of protection against electric shock: 

— Class II devices 

Classification by degree of protection against electric shock: 

— Applied part type B 

Classification by sterilization or disinfection method allowed by the manufacturer: 

— Refer to Sterilization. 

Classification by mode of operation: 

— Continuously operating device 

Thank you for purchasing the ENDO-MATE TC2.

Read this Operation Manual carefully before use for operation instructions and care and 

maintenance guidelines. Keep this Operation Manual for future reference.

Application

This product is a cordless handpiece used primarily for root canal enlargement. 

User

Only qualified personal is allowed to use the unit only in dentistry.

Prohibition

Do not use this motor handpiece for formation of an extremely bent root canal. Do not use this 

for implants other than endodontic treatment or other dental treatment.

Guidance and manufacturer's declaration - electromagnetic emissions 

The ENDO-MATE TC2 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the ENDO-MATE TC2 should assure that is used in such an environment. 

Emissions test 

Compliance

Electromagnetic environment - guidance

RF emissions
CISPR11

RF emmissions
CISPR11

Harmonic emissions
IEC61000-3-2

Voltage fluctuations/ 
flicker emissions
IEC61000-3-3

Group 1 
 
 
 

 
class  B 
 

 
class A 
 

 
Complies 
 

The ENDO-MATE TC2 uses RF energy only for 
its internal function. Therefore, its RF emissions 
are very low and are not likely to cause any 
interference in nearby electronic equipment.

The ENDO-MATE TC2 is suitable for use in 
all establishments, including domestic 
estabilishments and those directly connected 
to the public low-voltage power supply 
network that supplynetwork that spplies 
buidings used for domestic purposes.

Содержание Endo-Mate TC2

Страница 1: ...se read this Operation Manual carefully before use and file for future reference OM E0390E 001 Cordless Handpiece The EU directive 93 42 EEC was applied in the design and production of this medical de...

Страница 2: ...OPERATION MANUAL Cordless Handpiece English...

Страница 3: ...emely bent root canal Do not use this for implants other than endodontic treatment or other dental treatment Guidance and manufacturer s declaration electromagnetic emissions The ENDO MATE TC2 is inte...

Страница 4: ...unity The ENDO MATE TC2 is intended for use in the electromagnetic environment specified below The customer or the user of the ENDO MATE TC2 should assure that it is used in such an environment NOTE 1...

Страница 5: ...aximum output power rating of the transmitter in watts W according to the transmitter manufacturer Rated maximum output power of transmitter W 0 01 800MHz to 2 5GHz d 2 3 P 80MHz to 800MHz d 1 2 P 0 1...

Страница 6: ...iece into the charger check that the alarm sounds and the LED indicate charging animation for remaining battery capacity Unless charging is indicated this function is not performed and burns may resul...

Страница 7: ...Nationally Recognized Testing Laboratory NRTL in the United States and is accredited by the Standards Council of Canada to certify electro medical products with Canadian National Standards Follow the...

Страница 8: ...s can be set and program settings can be memorized by pressing this key for one second or longer The following parameters can be set Speed torque gear ratio and auto reverse setting UP DOWN Key Use th...

Страница 9: ...ies are being charged or in the refresh mode Full charge or nearly full charge About 30 80 remains Less than about 30 remains Batteries are drained or in a remarkably low voltage Charge the batteries...

Страница 10: ...inserted into the jack or the power switch is pressed with unnecessary force the cord or switch may be broken or a short circuit may occur Never use the battery charger for anything other than the mo...

Страница 11: ...battery charge goes down the actual load may not reach the preset torque limit value In this case this auto reverse function will not be activated If a load is continuously applied to the motor handpi...

Страница 12: ...t the time of factory shipment if setup becomes confusing 1 Turn off the power when the power is on 2 Hold down the PROGRAM key and ON OFF key for more than one second while pressing the POWER key at...

Страница 13: ...w the F type spray nozzle onto the Pana Spray nozzle by approx 10 turns 2 Insert the F type spray nozzle into the rear part of the contra angle head and lubricate the head for 2 3 seconds If the contr...

Страница 14: ...be at the end of their life expectancy In such a case ask your dealer to replace the batteries or replace them with new ones yourself Refer to the Changing Batteries When replacing them by yourself be...

Страница 15: ...atteries Outside the range of working temperature Abnormal heat generation from batteries Beyond the upper limit Below the lower limit Error Error code During rotation of the motor handpiece At the ti...

Страница 16: ...batteries Charge the batteries Use in a warm room Clean the contra angle head If nothing is displayed on the liquid crystal panel of the handpiece even when it is set to the charger failure in the cir...

Отзывы: