Liquid crystal panel 

ON/OFF key 

PROGRAM key 

DOWN key 

POWER key 

UP key 

SELECT key 

<Operation Panel> 

English

16 

17 

(4) Preparatory Operations 

1. Hold down the POWER key for more than one second 

to turn on the power. 

2. Keep holding down the PROGRAM key until program 

No. suitable for file to be used is displayed. 

3. When changing the set value of speed, torque, gear 

ratio or auto reverse, press the SELECT key, select the 

setting item and press the UP/DOWN key to change. 

(5) Operation 

If you press the ON/OFF key briefly, the motor handpiece 

starts. If you re-press the key, it stops. (Alternate 

operation) 

If you hold down the ON/OFF key for more than one 

second, the motor handpiece starts while the key is 

pressed. If you release the key, it stops. (Intermittent 

operation) 

 

    Auto Reverse Function 

Auto reverse setting can be changed by pressing the SELECT key for one second or longer 

while the motor handpiece stops. As the auto reverse mark flashes while the SELECT key is 

kept pressed, press the UP/DOWN key to adjust it. 

 

: Auto Reverse 

 

: Auto Stop 

 

: Auto Reverse OFF (flashes while the SELECT key is kept pressed, and goes out 

when it is released)  

 

The alarm will sound when it becomes a value of half of a torque limit set up during motor 

handpiece rotation, and the alarm will change near the torque limit value. (Load alarm) When it 

continues applying a load and it exceeds the torque limit value you can select mode from the 

following three modes. 

 

 

• When the battery charge goes down, the actual load may not reach the preset torque 

limit value. In this case this auto reverse function will not be activated. 

• If a load is continuously applied to the motor handpiece, it may automatically stop to 

prevent overheating. In this case, leave the motor handpiece for a while until it cools 

down. 

• If the ambient temperature is low, the alarm may sound when the motor handpiece 

rotates, however, this is not a failure. If the alarm does not stop sounding even when 

the ambient temperature increases, clean the contra angle head and perform 

calibration. (Refer to 6 (3) Calibration) 

CAUTION 

The GEAR RATIO key is fast-forwarded, when it is 

continuously held down. 

NOTICE 

Reverse rotation  
when load continues,  
forward rotation  
when load is removed 

Load lower than 

the set torque limit value 

Load higher than

the set torque limit value

 

AUTO REVERSING 

 

The handpiece stop that it reaches for the loaded value, the file rotates in reverse. When a load 

is removed, the file returns to normal rotation (forward rotation) automatically 

Reverse rotation 
when load continues,  
stop when load is removed 

Load lower than 

the set torque limit value 

Load higher than

the set torque limit value

Forward rotation

Returns to forward 

Reverse rotation  

Forward rotation 

Stop 

Reverse rotation 

Fig. 7

AUTO STOP

The motor handpiece starts in reverse. When a load is removed, the motor handpiece stops. If 

you want it to rotate (forward-rotate) again, re-press the ON/OFF Key or re-step.

 

AUTO REVERSE OFF 

 

The motor handpiece stops without reverse rotation. If you want it to rotate (forward-rotate) 

again, re-press the ON/OFF Key or re-step. 

Fig. 8

Fig. 6

Содержание Endo-Mate TC2

Страница 1: ...se read this Operation Manual carefully before use and file for future reference OM E0390E 001 Cordless Handpiece The EU directive 93 42 EEC was applied in the design and production of this medical de...

Страница 2: ...OPERATION MANUAL Cordless Handpiece English...

Страница 3: ...emely bent root canal Do not use this for implants other than endodontic treatment or other dental treatment Guidance and manufacturer s declaration electromagnetic emissions The ENDO MATE TC2 is inte...

Страница 4: ...unity The ENDO MATE TC2 is intended for use in the electromagnetic environment specified below The customer or the user of the ENDO MATE TC2 should assure that it is used in such an environment NOTE 1...

Страница 5: ...aximum output power rating of the transmitter in watts W according to the transmitter manufacturer Rated maximum output power of transmitter W 0 01 800MHz to 2 5GHz d 2 3 P 80MHz to 800MHz d 1 2 P 0 1...

Страница 6: ...iece into the charger check that the alarm sounds and the LED indicate charging animation for remaining battery capacity Unless charging is indicated this function is not performed and burns may resul...

Страница 7: ...Nationally Recognized Testing Laboratory NRTL in the United States and is accredited by the Standards Council of Canada to certify electro medical products with Canadian National Standards Follow the...

Страница 8: ...s can be set and program settings can be memorized by pressing this key for one second or longer The following parameters can be set Speed torque gear ratio and auto reverse setting UP DOWN Key Use th...

Страница 9: ...ies are being charged or in the refresh mode Full charge or nearly full charge About 30 80 remains Less than about 30 remains Batteries are drained or in a remarkably low voltage Charge the batteries...

Страница 10: ...inserted into the jack or the power switch is pressed with unnecessary force the cord or switch may be broken or a short circuit may occur Never use the battery charger for anything other than the mo...

Страница 11: ...battery charge goes down the actual load may not reach the preset torque limit value In this case this auto reverse function will not be activated If a load is continuously applied to the motor handpi...

Страница 12: ...t the time of factory shipment if setup becomes confusing 1 Turn off the power when the power is on 2 Hold down the PROGRAM key and ON OFF key for more than one second while pressing the POWER key at...

Страница 13: ...w the F type spray nozzle onto the Pana Spray nozzle by approx 10 turns 2 Insert the F type spray nozzle into the rear part of the contra angle head and lubricate the head for 2 3 seconds If the contr...

Страница 14: ...be at the end of their life expectancy In such a case ask your dealer to replace the batteries or replace them with new ones yourself Refer to the Changing Batteries When replacing them by yourself be...

Страница 15: ...atteries Outside the range of working temperature Abnormal heat generation from batteries Beyond the upper limit Below the lower limit Error Error code During rotation of the motor handpiece At the ti...

Страница 16: ...batteries Charge the batteries Use in a warm room Clean the contra angle head If nothing is displayed on the liquid crystal panel of the handpiece even when it is set to the charger failure in the cir...