NOVOTEST UT-3K-EMA Скачать руководство пользователя страница 3

 

 

Operating manual 

EMA Thickness Gauge NOVOTEST UT-3K-EMA 

RDC.ED.UT-3K-EMA.000 OM 

Page 3 

Caution! 

 

Please read this manual carefully before using EMA Thickness Gauge NOVOTEST UT-3K-EMA. 
 

Operating  manual  (hereinafter  referred  to  as  OM)  includes  the  information  required  to  inform 

operator  about  the  performance  and  operating  procedure  of  the  device  –  EMA  Thickness  Gauge 
NOVOTEST  UT-3K-EMA  (hereinafter  referred  to  as  device  or  thickness  gauge).  The  document 
contains the specification, description of the design and operating principle, as well as the information 
necessary for the right product operation. Before using the device,  you must read this manual, as  the 
device must be operated by persons acquainted with the operating principle design of the device. 

The right and efficient use of the device requires:  

 

measurement technique; 

 

conditions of measurement that correspond to the measurement technique;  

 

a trained operator who read this operating manual.  

NOVOTEST  reserves  the  right  to  make  minor  changes  that  do  not  impair  the  technical 

specification of the product. These changes may not be mentioned in this document. 

The delivery set of the device includes the operation documents that are a part of this manual and 

the device registration certificate. 

This OM applies to all product modifications. 

Содержание UT-3K-EMA

Страница 1: ...ddress and registered address 51200 Novomoskovsk Spasskaya Str 5 Sales office landline 38 056 767 23 25 multichannel mobile 38 067 565 82 33 sales novotest com ua http novotest ua Warranty repair and...

Страница 2: ...t 7 2 4 2 Selection of the amount of coherent integration 7 2 4 3 Specifics of the operation 8 2 4 4 Replacement of the probe 8 3 Safety precautions and warranty 8 3 1 Safety precautions 8 3 2 Verific...

Страница 3: ...ormation necessary for the right product operation Before using the device you must read this manual as the device must be operated by persons acquainted with the operating principle design of the dev...

Страница 4: ...e tab 1 1 Table 1 1 Principal specifications of the device Measurement range for steel mm 2 60 Thickness measurement error mm 0 08 The permissible gap between the probe and the test object mm max 2 Th...

Страница 5: ...measured time is recalculated into the thickness value by the preset ultrasound velocity value The acoustic wave is formed by electromagnetic acoustic transduction directly on the surface of the test...

Страница 6: ...st be charged via the charger If the battery charge is less than 25 the device readings will flash on and off When the charger is connected to the device the display shows the level of the battery cha...

Страница 7: ...randomly 2 Calibrate the device if it is necessary according to c 2 3 3 Place the device on the surface of the test object The device will display the thickness value within a period from 100 millisec...

Страница 8: ...and hold the device meanwhile 4 It is recommended to put the device on the test object at a 60 angle Once the probe touches the test object the device should be put upright Throughout the device oper...

Страница 9: ...tive air humidity max 80 at the temperature 25 air pressure from 86 kPa to 106 kPa 3 2 2 Steps of the verification and accessories When performing the verification take the following steps and use acc...

Страница 10: ...10 and 60 mm reference blocks Verification is considered acceptable if the measurement error of the device for all reference blocks is not more than 0 04 mm 3 2 6 Execution of verification The verifi...

Страница 11: ...ar during the device operation However if these parts fail during the warranty period due to a defect in material or workmanship they will be repaired or replaced under the basic warranty 3 3 4 Owner...

Страница 12: ...ipping expenses the device loss of profit or damage to property all warranties shall become void if the device is officially recognized unserviceable 3 3 7 Warranties and consumer law The basic warran...

Страница 13: ...kness Gauge NOVOTEST UT 3K EMA RDC ED UT 3K EMA 000 OM Page 13 If any failure in the device is found it shall be submitted to NOVOTEST for maintenance The table 3 3 lists the failures that operator ca...

Страница 14: ...vidence Possible reason Method of elimination No digits on the display after turning on The battery is fully discharged Charge the battery When checking the performance of the thickness gauge on a ref...