Damages  caused  by  negligent/incorrect  operation  of  the  device,  natural  disaster,  water 
intrusion into device, probe, accessories, and parts of the device, accident or misuse of the 
device; 

 

Wear and tear of parts; 

 

Non-observance  of  NOVOTEST  recommendations  regarding  periodicity  of  the  device 
maintenance; 

 

Non-observance of requirements to the device operation recommended by NOVOTEST; 

 

Changing of the configuration of the device or its components,  intrusion into systems of 
the device, etc. without the approval of NOVOTEST; 

 

Use of inappropriate batteries and accessories; 

 

Changing of power supply in the grid; 

 

Refusal from repair of any damages found during the regular maintenance; 

 

Factors outside the scope of control of NOVOTEST, for example, air pollution, hurricanes, 
chips resulting from shocks, scratches, and use of unsuitable cleaning fluids; 

 

Use of the repair methods not approved by NOVOTEST; 

 

Use of third-party spare parts and couplants of NOVOTEST. 

All  repairs  covered  by  the  NOVOTEST  warranty  shall  be  performed  by  the  authorized  service 

center of NOVOTEST. 

 

3.3.6 Cases uncovered by warranty 

The  basic  warranty  of  NOVOTEST  and  extended  warranty  of  NOVOTEST  shall  not  hold 

NOVOTEST  accountable  for  any  unexpected  or  indirect  damage  resulting  from  a  defect  covered  by 
the above warranties. Such damage includes (but is not limited to the following): 

 

reimbursement for inconveniences, phone calls, posting and shipping expenses the device, 
loss of profit or damage to property; 

 

all warranties shall become void if the device is officially recognized unserviceable. 

 

3.3.7 Warranties and consumer law 

The  basic  warranty  of  NOVOTEST  and  extended  warranty  of  NOVOTEST  shall  not  harm  your 

rights provided by the sale agreement that is  executed with  selling  the device to  the manufacturer or 
authorized  dealer  of  NOVOTEST;  and  the  applicable  law  stipulating  the  requirements  to  sales  and 
servicing of consumer goods. 

 

3.4 Maintenance 

The information about the maintenance applies to all NOVOTEST products. 
The maintenance shall be performed during the device operation and is categorized into: 

 

preventive maintenance; 

 

regular maintenance. 

Preventive  maintenance  shall  be  performed  at  least  once  in  a  quarter  and  includes  visual 

inspection, cleaning, and greasing. 

Regular maintenance shall be performed by NOVOTEST at least once a year and is mandatory to 

save the warranty of NOVOTEST. 

It is very important to maintain the device during the entire operation period. 
The  list  of  steps  performed  during  each  maintenance  will  depend  on  the  device  model,  year  of 

production,  and  run  to  failure.  By  your  request,  the  authorized  service  center  of  NOVOTEST  may 
provide you the information about steps to be taken when servicing the device. 

The records of regular maintenance of your device shall be made in the registration certificate of 

the  device.  The  information  about  maintenance  is  very  important;  it  can  be  needed  to  exercise  your 
rights  for  warranty  repair  of  the  device.  So,  always  make  sure  that  the  authorized  service  center  of 
NOVOTEST puts a stamp in the right place under the record about the performed procedure. 

Содержание UT-3K-EMA

Страница 1: ...ddress and registered address 51200 Novomoskovsk Spasskaya Str 5 Sales office landline 38 056 767 23 25 multichannel mobile 38 067 565 82 33 sales novotest com ua http novotest ua Warranty repair and...

Страница 2: ...t 7 2 4 2 Selection of the amount of coherent integration 7 2 4 3 Specifics of the operation 8 2 4 4 Replacement of the probe 8 3 Safety precautions and warranty 8 3 1 Safety precautions 8 3 2 Verific...

Страница 3: ...ormation necessary for the right product operation Before using the device you must read this manual as the device must be operated by persons acquainted with the operating principle design of the dev...

Страница 4: ...e tab 1 1 Table 1 1 Principal specifications of the device Measurement range for steel mm 2 60 Thickness measurement error mm 0 08 The permissible gap between the probe and the test object mm max 2 Th...

Страница 5: ...measured time is recalculated into the thickness value by the preset ultrasound velocity value The acoustic wave is formed by electromagnetic acoustic transduction directly on the surface of the test...

Страница 6: ...st be charged via the charger If the battery charge is less than 25 the device readings will flash on and off When the charger is connected to the device the display shows the level of the battery cha...

Страница 7: ...randomly 2 Calibrate the device if it is necessary according to c 2 3 3 Place the device on the surface of the test object The device will display the thickness value within a period from 100 millisec...

Страница 8: ...and hold the device meanwhile 4 It is recommended to put the device on the test object at a 60 angle Once the probe touches the test object the device should be put upright Throughout the device oper...

Страница 9: ...tive air humidity max 80 at the temperature 25 air pressure from 86 kPa to 106 kPa 3 2 2 Steps of the verification and accessories When performing the verification take the following steps and use acc...

Страница 10: ...10 and 60 mm reference blocks Verification is considered acceptable if the measurement error of the device for all reference blocks is not more than 0 04 mm 3 2 6 Execution of verification The verifi...

Страница 11: ...ar during the device operation However if these parts fail during the warranty period due to a defect in material or workmanship they will be repaired or replaced under the basic warranty 3 3 4 Owner...

Страница 12: ...ipping expenses the device loss of profit or damage to property all warranties shall become void if the device is officially recognized unserviceable 3 3 7 Warranties and consumer law The basic warran...

Страница 13: ...kness Gauge NOVOTEST UT 3K EMA RDC ED UT 3K EMA 000 OM Page 13 If any failure in the device is found it shall be submitted to NOVOTEST for maintenance The table 3 3 lists the failures that operator ca...

Страница 14: ...vidence Possible reason Method of elimination No digits on the display after turning on The battery is fully discharged Charge the battery When checking the performance of the thickness gauge on a ref...

Страница 15: ...ght The device shall not be exposed to rapid shocks falls or strong vibrations The devices shall be placed on storage racks or in palletized in the transportation container For a long term storage the...

Страница 16: ..._______________________________________________________________ __________________________________________________________________________________ _____________________________________________________...

Отзывы: