NOVOTEST.UT-1.000 OM 

Operation manual 

Ultrasonic Thickness Gauge NOVOTEST UT-1M 

 

Page. 14 

6.

 

Rename the changed calibration by pressing the key  «

» open the action menu with the 

calibrations, highlight the "MOVE" item and confirm with the key «

». 

7.

 

Enter  a  new  calibration  name.  Moving  the  cursor  on  the  virtual  keyboard  is  done  using  the 

keys  «

» and «

», and confirm with key «

». To save a new name, you need 

to move the cursor to the "SAVE" button and press the enter key «

». 

 

2.3.5 Work in "INFORMATION" mode 

In  this  mode,  you  can  find  data  about  the  manufacturer.    Also  in  the  "INFORMATION"  mode, 

you can see the information about the device and the number of measurements. For this, using keys «

» and «

» find the needed page. 

 

2.4 Making measurements 

Before  using  the  device  for  measurements,  make  sure  that  the  battery  has  a  sufficient  level  of 

charge; the level of charge is indicated on the graphical display in the form of a battery (the indicator). 
A completely dark indicator indicates that the battery is  100% charged. One line inside the indicator 
field corresponds to approximately 20% of the charge volume.  

If there is no or insufficient charge, recharge the battery using charger. 
 

2.4.1 Preliminary calibration of the probe 

To  exclude  the  influence  of  the  transit  time  of  ultrasonic  oscillations  in  the  delay  lines  on  the 

result of measuring the thickness  of the product,  it is  necessary to  calibrate the connected transducer 
before making measurements, for which purpose: 

1.

 

In the "MEASUREMENT" mode, press the key «

» And select the "CALIBER" mode. 

2.

 

Set using the keys  «

» and «

»The propagation velocity of the US, corresponding 

to the reference measure on the device (6070 m/s). 

3.

 

Clean the contact surface of the inverter from grease, if any. 

4.

 

Put the transducer into the air, according to the instruction, on the display of the device "PULL 

IN  AIR",  then  confirm  the  action  by  pressing  the  key  «

».  The  instrument  will  take  a 

measure and display a new instruction. 

5.

 

Apply a layer of contact lubricant on the built-in reference thickness sample block. 

6.

 

Install the transducer’s  contact surface on the reference measure and press firmly, according 
to the instructions, on the displays of the device "INSTALL ON THE SAMPLE", then confirm 

the action by pressing the key «

». The instrument will measure and after the display of 

the results go into the normal measurement mode. 

 

Caution! 

 

Preliminary calibration of the transducer should be carried out only on the sample (measure)  built-in 
the thickness gauge. 
 

The operation of preliminary calibration of the transducer is desirable to be carried out in the event 

of  changes  in  operating  conditions  (a  significant  change  in  the  ambient  temperature),  when  the 
converter is first turned on, after a prolonged intensive operation of the converter. 

 

Содержание UT-1M

Страница 1: ...ULTRASONIC THICKNESS GAUGE NOVOTEST UT 1M Operating Manual 2017 ...

Страница 2: ...1 9 Packagin 8 2 Intended use 9 2 1 Operational limitations 9 2 2 Preparing for use 9 2 2 1 Visual inspection 9 2 2 2 Installing the batteries 9 2 2 3 Connectiong the probe 9 2 3 Using the device 10 2 3 1 Switching on 10 2 3 2 Operation in MEASUREMENT mode 11 2 3 3 Operation in the SETTINGS mode 12 2 3 4 Operation in CALIBRATION mode 13 2 3 5 Operation in INFORMATION mode 14 2 4 Making measurement...

Страница 3: ... actions as well as information necessary for the correct use of the products Before use it is necessary to study this document as it is to be used by those familiar with the principles of operation and the design of the device Correct and effective use of the control device requires mandatory availability methods of control conditions of control incoming control methods trained and studied user m...

Страница 4: ...rmation processing unit max mm 122х65х23 Overall dimensions of the converter no more than mm 25х40 Powered by two NiMH batteries or AA batteries 1 2 V Time of continuous operation not less than h 8 Weight of electronic unit with battery less than kg 0 25 Transducer weight less than kg 0 15 Operating temperature range of the information processing unit С from 10 to 40 Operating air humidity less th...

Страница 5: ...sented in Table 1 2 Table 1 2 Metrological characteristics of converters Transducer type Measurements range Measurement error P112 10 6 2 0 8 30 mm 0 01h 0 05 mm P 112 5 10 2 1 75 mm 0 01h 0 05 mm P 112 2 5 12 2 2 5 300 mm 0 01h 0 05 mm P 112 1 25 20 2 4 300 mm 0 01h 0 05 mm where h nominal thickness mm Additional time error for 2 hours of continuous operation is no more than the basic 1 5 Parts T...

Страница 6: ...ge 9 protective silicone bumper 10 transducer 11 generator plug 12 battery compartment 13 receiver plug 14 battery compartment cover Figure 1 1 Ultrasonic thickness gauge NOVOTEST UT 1M Figure 1 2 Keyboard Buttons and Their Functions где power key Left soft key right soft key move up in the instrument menu increase the input values move down to the instrument menu decrease input values measurement...

Страница 7: ...e the correct propagation velocity of US in this particular material The actual speed of ultrasound in different materials is very different both from material to material and from the reference data The best results can be obtained if the instrument is previously calibrated using an object of a known thickness reference made of the same material as the object under study The standard should have ...

Страница 8: ...or use 2 2 1 Visual inspection Inspect the device ensure that there is no mechanical damage to the information processing unit transducer cable plugs and sockets 2 2 2 Installing the batteries Insert the batteries into the battery compartment by opening the battery cover lightly and pulling it down until the cover is completely detached The batteries are to be installed according to the polarity i...

Страница 9: ...n on Fig 2 3 Caution When disconnecting the plug from the socket grasp its housing in a corrugated area and in no case should you pull the cable Figure 2 3 Manipulating the LEMO connectors 2 3 Using the device 2 3 1 Switching on Switch the device on by a long press key on the control panel until a short time splash screen appears Figure 2 4 on the graphic display Figure 2 4 the splash screen After...

Страница 10: ...asured values DELETE ALL go to the main menu EXIT and also cancel the press and close the CANCEL selection menu 2 3 2 1 Measuring modes The device operates in several measurement modes to switch between modes use keys After it is pressed the menu for selecting the mode of operation appears NORM STAT B SCAN and CALIBER The NORM mode is intended for carrying out the usual measurement without outputt...

Страница 11: ...h Russian etc BRIGHTNESS change the brightness of the display from 0 to 100 the resolution is 10 AUTO OFF Set the device to automatically turn off when not in use off 1 5 10 30 minutes RETRO allows you to return to the measurement mode with the last measurements saved after the device is rebooted SCALE select the units of measurement mm or inch inches MATERIAL selection of the material of the obje...

Страница 12: ... keys and select the calibration for the inverter to be used 2 2 By pressing the key open the action menu with the calibrations highlight the item SELECT and confirm with the key 2 3 4 2 Creating a new inverter calibration To create a new calibration you need to 1 Using keys and select a saved calibration with the name PROBE for editing 2 Using key open the action menu with the calibrations highli...

Страница 13: ...ns in the delay lines on the result of measuring the thickness of the product it is necessary to calibrate the connected transducer before making measurements for which purpose 1 In the MEASUREMENT mode press the key And select the CALIBER mode 2 Set using the keys and The propagation velocity of the US corresponding to the reference measure on the device 6070 m s 3 Clean the contact surface of th...

Страница 14: ...owever it should be borne in mind that the thickness of the paint will go into the resulting measurement result 2 4 3 Measurements To make measurements 1 Set the speed of propagation of ultrasound in the object of control To do this go to the SETUP mode and select the material of the monitoring object in the MATERIAL parameter If the required material is not in the base of the device you need to s...

Страница 15: ... transport 4 4 During transportation handling and storage of stock the device should be kept from shock bumps and moisture 5 MAINTENANCE 5 1 Maintenance of the device is made by the manufacturer in the event of faults in the device operation 6 WARRANTY 6 1 The manufacturer guarantees the normal operation of the unit for 12 months from date of purchase and undertakes to make repair it during this t...

Страница 16: ..._________________________________________________________________ __________________________________________________________________________________ __________________________________________________________________________________ __________________________________________________________________________________ __________________________________________________________________________________ __...

Отзывы: