The "CALIBER" mode is intended for the pre-measurement calibration of the converter (zero-
setting of the converter) before the start of the measurement, in order to exclude the influence of the
transit time of ultrasonic oscillations in the delay lines.
Also in the device there is a mode of calibration of amplification of a signal by the receiver, for
transition in this mode it is necessary to press a key «
», After that, the speed of the ultrasound in
the material changes to the gain adjustment in dB (Figure 2.9). To change the gain use the keys «
» и «
». Для выхода с режима калибровки усиления нужно повторно нажать клавишу
«
».
Figure 2.9 - Gain calibration mode
2.3.3 Working in the "Settings" mode
When the "SETTINGS" menu item is selected, the device enters the setting mode of the following
parameters (Figure 2.10):
"LANGUAGE": selection of the menu language of the device (available in English,
Russian, etc);
"BRIGHTNESS": change the brightness of the display (from 0 to 100%, the resolution is
10%);
"AUTO OFF": Set the device to automatically turn off when not in use (off, 1, 5, 10, 30
minutes);
"RETRO": allows you to return to the measurement mode with the last measurements
saved after the device is rebooted;
"SCALE": select the units of measurement (mm or inch (inches));
"MATERIAL": selection of the material of the object to be monitored (User; Steel 40X13;
Steel St3; Steel X15N15GS; Steel XN35VT; Steel XN70BMTY; Steel 40HNMA; Steel
XN77UYUR; Steel 20GNDM; Bronze; Iron; Tungsten; Aluminum; D16T; Aluminum;
Silver; Gold; Lead, Tin, Nickel, Molybdenum, Copper, Manganese, Magnesium, Brass
LS-51-1, Brass, Capron, Caprolon, ED-5, Ebonite, Cast Iron, Zinc, Chrome, Porcelain,
Teflon, Textolite, Silicate glass; Org Glass, Rubber, Polystyrene, Plexiglas);
"REFERENCE": thickness of the reference sample (mm);
"SPEED REF": speed of propagation of ultrasound in the reference sample (m / s);
"TOP": upper measuring range (mm);
"TOP": upper measuring range (mm).
Figure 2.10 - "SETTINGS" mode
To make changes in the "SETTINGS" mode, you need to select the required parameter with the
keys «
» and «
», then confirm by pressing «
» and change the setting to the desired