7.3. The battery status 
3 segments indicate battery status. The larger number of  displayed 
segments, the more energy remains in the battery. When low energy 
level,  the  battery  symbol  starts  to  blink,  and  the  battery  should  be 
charged.  Connector  for  charging  is  located  in  the  upper  end  of  the 
unit. 
 

8. Operating  

The  device  measures  the  three  different  values:  air  temperature, 
surface  temperature and  relative  humidity.  From  these  values  shall 
be calculated dew point temperature and 

ΔT.  

 
8.1. Relative humidity 
For  the  accurate  measurement  of  relative  humidity  is  necessary  to 
ensure unimpeded air flow through the grooves of the sensor. 
 
8.2. Surface temperature  
The  sensor for measuring  the  surface  temperature has  a  metal  tip. 
Metal tip is placed against the surface, preferably at an angle of 90 
°,  and  fixed  in  this  position  until  the  stabilization  of  the  measured 
values.  During  measuring  u

ser  needn’t press  on  the  Device  with  a 

great force. 
 
8.3. Air temperature 
The  sensor  measures  the  temperature  of  the  air  directly  in  their 
environment, and so it is important to keep items which could affect 
on the temperature of the air, for example, the fingers away from this 
sensor. 
 
8.4. Dew point temperature 
Dew  point  temperature  is  calculated  parameter  which  is  obtained 
from the values of air temperature and relative humidity.  

Содержание KTR-1

Страница 1: ...Dew Point Meter NOVOTEST KTR 1 OPERATING MANUAL 2015...

Страница 2: ...nd system Dew Point Meter NOVOTEST KTR 1 2 CONTENTS 1 Introduction 3 2 Appointment 3 3 Technical characteristics 3 4 Completeness 4 5 Labeling packaging 4 6 Safety precautions and maintenance 5 7 Desi...

Страница 3: ...culation and measurement the surface temperature by contact method 2 2 Dew Point Meter NOVOTEST KTR 1 allows user to measure air temperature 20 to 125 C surface temperature 20 to 125 C relative humidi...

Страница 4: ...ace temperature 1 Calculated values dew point temperature 15 40 difference between dew point and surface temperature Depends on measurements Dimensions mm 96 47 24 Operating temperature range 10 40 Po...

Страница 5: ...heavy loads and extreme temperatures Overheating can damage the screen plastic housing and internal parts Dust and prolonged high humidity can damage the unit causing malfunction The device is not wa...

Страница 6: ...KTR 1 6 Press and hold leads to on off the device Control of button illumination Long pressing leads to overturning on 180 degrees the information at the display On off of displaying of all measured p...

Страница 7: ...ntrol units and system Dew Point Meter NOVOTEST KTR 1 7 Surface temperature measurement mode Relative humidity measurement mode Dew point temperature displaying mode Displaying mode of difference betw...

Страница 8: ...For the accurate measurement of relative humidity is necessary to ensure unimpeded air flow through the grooves of the sensor 8 2 Surface temperature The sensor for measuring the surface temperature h...

Страница 9: ...r quality painting due to the formation of condensation on the surface A warning signal sounds only in the mode of displaying difference between the surface temperature and the temperature at point 9...

Страница 10: ...10 10 Warranty 10 1 The manufacturer guarantees compliance device to the technical requirements for observance of consumer terms of transportation storage and operating 10 2 Warranty period 12 months...

Отзывы: