background image

 

Development and manufacturing of 

non-destructive control units and 

system 

 

Dew Point Meter NOVOTEST KTR-1 

9

 

 

Surfaces  with  surface  temperature  near  or  below  the  dew  point 
temperature will be covered by condensate.  
 
8.5

. ΔТ  

ΔT  -  is  the  difference  between  the  surface  temperature  of  tested 
objects and the dew point temperature.  
 
8.6. The warning signal 
Devices  makes  sound  alarm  when 

ΔT  (difference  between  surface 

temperature  and  dew  point) 

decreases  more  than  3  °  C.  In  such 

climatic conditions there is a risk of poor-quality painting, due to the 
formation of condensation on the surface. A warning signal sounds 
only  in  the  mode  of  displaying  difference  between  the  surface 
temperature and the temperature at point.  

 
9. Storage and transportation 

9.1. The Device must be stored at ambient temperature from  -10 to 
+35 

° C and relative humidity up to 80% at 25 ° C. 

 
9.2.  The  storage  should  be  free  of  dust,  fumes,  acids,  alkalis  and 
corrosive gases. 
 
9.3. Transporting device in a package can be made by any mode of 
transport  in  accordance  to  the  requirements  and  rules  of 
transportation, operating on the data types of transport. 
 
9.4. During transportation, handling and storage of stock unit should 
be kept from blows, shocks and moisture. 

 

 
 
 

Содержание KTR-1

Страница 1: ...Dew Point Meter NOVOTEST KTR 1 OPERATING MANUAL 2015...

Страница 2: ...nd system Dew Point Meter NOVOTEST KTR 1 2 CONTENTS 1 Introduction 3 2 Appointment 3 3 Technical characteristics 3 4 Completeness 4 5 Labeling packaging 4 6 Safety precautions and maintenance 5 7 Desi...

Страница 3: ...culation and measurement the surface temperature by contact method 2 2 Dew Point Meter NOVOTEST KTR 1 allows user to measure air temperature 20 to 125 C surface temperature 20 to 125 C relative humidi...

Страница 4: ...ace temperature 1 Calculated values dew point temperature 15 40 difference between dew point and surface temperature Depends on measurements Dimensions mm 96 47 24 Operating temperature range 10 40 Po...

Страница 5: ...heavy loads and extreme temperatures Overheating can damage the screen plastic housing and internal parts Dust and prolonged high humidity can damage the unit causing malfunction The device is not wa...

Страница 6: ...KTR 1 6 Press and hold leads to on off the device Control of button illumination Long pressing leads to overturning on 180 degrees the information at the display On off of displaying of all measured p...

Страница 7: ...ntrol units and system Dew Point Meter NOVOTEST KTR 1 7 Surface temperature measurement mode Relative humidity measurement mode Dew point temperature displaying mode Displaying mode of difference betw...

Страница 8: ...For the accurate measurement of relative humidity is necessary to ensure unimpeded air flow through the grooves of the sensor 8 2 Surface temperature The sensor for measuring the surface temperature h...

Страница 9: ...r quality painting due to the formation of condensation on the surface A warning signal sounds only in the mode of displaying difference between the surface temperature and the temperature at point 9...

Страница 10: ...10 10 Warranty 10 1 The manufacturer guarantees compliance device to the technical requirements for observance of consumer terms of transportation storage and operating 10 2 Warranty period 12 months...

Отзывы: