Development and manufacturing of non-
destructive testing units and system
Ultrasonic Tester of Building Materials Strength NOVOTEST IPSM-U
23
the convenience of it s replacement when it is required or sharp
reduction of the continuous operation time independently of the
country. Replacement is possible only by the battery with similar
characteristics in compliance with the marking on it. From
environmental protection point of view, the best thing is to use the
battery.
9.5. Storage
9.5.1. Device shall be kept in the carry case, the probe and the
batterys shall be disconnected.
9.5.2. If device is kept in the carry case for than 14 days, the battery
shall be taken out from its compartment in the information
processing unit.
9.5.3. It is recommended to keep device in closed premises with the
relative humidity not more than 80%, there shall be no mold, paints,
acids, chemical agents and other chemicals, the evaporation of
which my give a harmful effect. Sharp fluctuations of temperature
and humidity which can result in dew formation are not allowed.
9.6. Transportation
9.6.1. Device transportation in the carry case shall be only in closed
vehicles, where the possibility of mechanical damage or
atmospheric precipitation is excluded.
9.6.2. The way packed in carry cases devices are located inside the
vehicle shall exclude.
9.7. Putting into operation after storage and transportation.
After storage or transportation under the temperature lower than -
5°C, before starting device operation, it is necessary to keep it not
longer than 1 hour under the temperature higher than +10°C and not
less than 2 hours under the temperature higher than 0°C.