Development and manufacturing of non-
destructive testing units and system
Ultrasonic Tester of Building Materials Strength NOVOTEST IPSM-U
13
8.
Operation
8.1. Preparation for use.
8.1.1. Connection of the probe.
Connect the receiver and the transmitter to the connectors on the
top of the electronic processing unit.
Attention! To prevent instrument failure and loss guarantee,
connection to the instrument cables of the probe must be doing
with the power off device, preventing accidental circuit of the
output device on one of its input connectors.
8.2. Battery operation
8.2.1. Install the battery into the battery compartment, observing the
proper polarity.
8.2.2. To control the battery charge, turn on the device by long
pressing the button
.
8.2.3. Battery level is shown on the top left of the display electronics
unit symbol "battery." Completely dark square inside the symbol
indicates a full charge the battery. The battery squares disappear
sequentially, from left to right. One dark square or squares absence
means need to recharge the battery. The "battery" is on display in
any mode of the device.
8.2.4. To charge the battery, turn off the device by long pressing the
button
, and then remove the battery from the battery
compartment and make its charge in accordance with p. 8.2.5.
8.2.5. Charging the battery
To charge the battery:
- connect the battery to the charger contacts;
- turn the charger in socket.