29
Pen approaching end of life
The battery is running low, and the dose
memory is approaching end of life.
What to do?
The pen will last four to five years. When the
battery is running low and the dose memory
display has shown “End” for a while, the display
will turn blank and stop working.
Contact your supplier to get a new pen. You
can still use your pen for insulin delivery until
you get a new one.
See page 32
for warranty
information.
Last injection not registered
The dose memory has not registered your last
injection.
What to do?
Normally, the dose memory will start working
again after your next injection.
If not, contact your supplier to get a new pen.
You can still use the pen for insulin delivery
without using the dose memory.
If you are in doubt about how much insulin you
injected, check your blood sugar level.