9
3
3B
3A
•
Press the dose button
until the dose counter shows “0”. The
“0” must line up with the dose pointer.
• Check that
insulin squirts
from the needle tip.
If no insulin appears,
repeat steps 1-3 until insulin squirts from
the needle tip.
If no insulin appears after several attempts, check the insulin
window to see if there is a gap between the black piston rod
head and the piston,
see illustrations 3A and 3B
. If you do not see
any gap, change the needle and test the insulin flow again.
The insulin flow test is complete when you see insulin
squirt from the needle tip.
Do not try to inject if no insulin appears.
You may not
receive any insulin, even though the dose counter may
move. This could lead to too high blood sugar
(hyperglycaemia).
If insulin does not appear after testing the flow
repeatedly,
do not use the pen. Contact your supplier to
get a new pen.