1. Introduction

1.1 You are awesome! 

Thank you for having purchased a Novasina LabMaster-aw neo water activity 

meter. You have opted-in for cutting-edge functionality paired with very precise, 

accurate and reliable sensor technology made in Switzerland. A fantastic symbi-

osys! This top of the line instrument allows you to make and get most out of a 

water activity measurement but ensure safe, reliable operation and full 21CFR11 

compatability, where this standard is applicable for such a benchtop meter. 

Please read these operating instructions carefully before setting the unit into 

operation and contact us if you feel unsure about a certain task or procedure. 

We are here to help you. Your local representative is listed on our website www.

novasina.com

1.2 Intended Use 

The Novasina LabMaster-aw neo is exclusively intended

for measuring WATER ACTIVITY under strict adherence to the information

and notes given in this manual. Any use beyond this scope is considered

to be a violation of the intended purpose and may endanger your safety

or could result in the product being damaged. Any use of this Novasina

instrument other than specified is at customer’s own risk. By any

means will the manufacturer or supplier be liable for any resulting damage. 

Safety Instructions

• 

The Novasina LabMaster-aw neo should be serviced, maintained

• 

A LabMaster-aw neo must not be used in explosive or hazardous zones. It is 

designed to be operated in laboratory environments.

• 

Do not measure explosive or highly flammable materials/samples with the 

LabMaster-aw neo. 

• 

Before connecting the unit to power, ensure that mains voltage complies 

with rated voltage and frequency of Novasina LabMaster-aw neo (check 

label on the back) which is 100...240V AC / 50..60 Hz. Verify as well that plug 

and cables do not have any visible damage.

• 

Das LabMaster-aw neo system may be used only under the specified opera-

ting conditions (see Spec Sheet, downloadable on our website)

• 

Please consider and respect local regulation in terms of the usage ond ope-

ration of electronic measurement instruments. 

• 

Please use original accessories and spare parts from your local representati-

ve only. Those parts are listed on our website, www.novasina.com.

• 

Without written permission of the manufacturer, it is forbidden to perform 

any technical change (mechanical, electronical, hardware and software) at 

this aw-meter.

• 

Power must be disconnected before the housing of LabMaster-aw neo is 

opened (disconnect power plug completely)

• 

The workplace must be set-up according current ESD regulations and a ser-

vice technician must use and wear all necessary ESD implements. Otherwise, 

defects on components cannot be excluded. 

• 

Unanthorized opening of the LabMaster-aw neo will lead to immediate loss 

of warranty

1.3 Protective Measures

Please take note and respect protective measures to avoid instrument problems 

due to improper handling. Warranty is gone in that case.

Содержание LabMaster-aw neo

Страница 1: ...aw Messger t aw measuring instrument Appareil de mesure aw Novasina versions V1 11 and higher Operating Instructions LabMaster aw neo Doc No 005739 04...

Страница 2: ...a Stable System 15 4 2 2 Selectable Stability Modes 16 4 2 3 Which mode for which products 16 4 2 4 Stable Indicator Progress bar 17 4 2 5 Stability check temperature 17 4 3 Measurement Temperature Be...

Страница 3: ...1 Reset a Forgotten or Lost Password 40 11 3 Administration of Permission Profiles 41 11 3 1 Adjustable Permissions within Permission Profiles 41 11 4 Administration of User Accounts 43 11 4 1 Adjust...

Страница 4: ...abMaster aw neo must not be used in explosive or hazardous zones It is designed to be operated in laboratory environments Do not measure explosive or highly flammable materials samples with the LabMas...

Страница 5: ...tic acid strong aroma and or essential oils All those volatiles affect a water activity measurement if any preventive measures are in place and can lead to sensor destruction in a very short time Nova...

Страница 6: ...es or license purchases needed If you open the box you should find 1x LabMaster aw neo 1x Carton box with 2x ePW dispenser with 20 sample cups each plus a dispenser support 1x USB and 1x RS232 cable 1...

Страница 7: ...r aw neo with pleasant anticipation now we instruct you how you can get a perfect start working with the new aw meter Got lost somewhere No worries there is a Help System on the instrument Simply pres...

Страница 8: ...easurement The gre ener the bar the closer to completion Stable mode is displayed underneath of the bar Method Name If a method is loaded from sample database method name is displayed at the bottom of...

Страница 9: ...the day are displayed in this area Press on the audit trail area to see the complete audit trail and to be able to filter certain events inclu ding last masurement results Every entry consist of the f...

Страница 10: ...amber Do not forget to cover sample cup or SAL T to avoid humidity exchange with air humidity from ambient air 3 1 6 Status Bar The following information is displayed alternating Text Symbol Descripti...

Страница 11: ...stable values in the stable display or graph area Press on this icon to go back to measurement screen This button is also available in submenus Accessing system settings to adjust system parame ters...

Страница 12: ...et tem perature for temperatur stability See chapter 4 2 5 This menu item is only dis played if Tempe rature control is on and Stable Mode is NOT set to Quick Off Graph Para meters Settings for graph...

Страница 13: ...s ISO AOAC etc or internal guidelines request a water acti vity measurement at a specific temperature i e 25 C The LabMaster aw neo allows a measurement temperature setting between 0 C 32 F and 60 C 1...

Страница 14: ...face temperature is kept constant to target value 0 02 C If instrument is operated in completely new conditions or target value is changed remarkably it can take up to 30mins until target value 0 30C...

Страница 15: ...in the air move towards an eqilibrium the net amount of water which is exchanged becomes smaller and smaller Therefo re the change in air humidity and water activity finally becomes smaller in the sam...

Страница 16: ...at line mode for production every measurement is finished within max 10 minutes as long as temperature is set in the range of 20 C to 30 C Query a query about the requested stability time is issued e...

Страница 17: ...difference bet ween sample surface IR and sensor If message persists open chamber check if measurement table and outer metal ring of measurement head are clean and free of dust and dirt and redo the m...

Страница 18: ...oblems can appear if target measurement temperature is pro grammed to a value below the dew point temperature Once the measurement chamber has opened small water droplets are for med within the measur...

Страница 19: ...e ceramics as it absorbs free water and can change the water activity value 2 Fill sample cup half to 3 4 Do not overfill as it will just squeeze sample into the protection filter which results in fil...

Страница 20: ...hanical protection filter which is installed by default must not be removed Make sure the chamber is absolutely dry If small water droplets are present it might lead to wrong verification calibration...

Страница 21: ...he measurement position and close the measu rement chamber A pop up window as below is displayed It contains info about the used salt standards as theoretical value at 25 C serial number last usage an...

Страница 22: ...t on the RFID reader in the measurement chamber indicated with RFID symbol until you hear a short beep then move the standard to the measurement position and close the measurement chamber A pop up win...

Страница 23: ...calibration point is only brought up to date once a calibration had been performed at that point In case of a verification sensor health will be simulated and displayed after the process but not stor...

Страница 24: ...calibrated points are listed Depending whether a printer or a PC is connected a pop up windows appears in which print destination can be selected Set sensor pass word Set a sensor password Important...

Страница 25: ...Plot of Measurement Sequence The LabMaster aw neo offers great opportunities to plot and analyze a measure ment sequence especially if a new product is tested and running in sequence is not very well...

Страница 26: ...Description Skip to the start of the graph Scroll backwards Scroll forwards Skip to the end of the graph If a black dot is visible graph is scrolled to the end automatically so that the latest values...

Страница 27: ...er of the records This field can be left empty if no name is applicable or necessary Sample number Serves as allocation of a protocol to a sample This sample number is displayed in the chamber area of...

Страница 28: ...tion Note Factory Settings Product name Name of Sample Will be stored in recorded data Sample number Sample number for next mea surement This number will be increased automatically as soon as a new or...

Страница 29: ...print the protocol to the SD card For an on going measurement only actual Protocol is available for a finished measurement a stable protocol can be printed 8 3 3 Connect a Local Printer It is possibl...

Страница 30: ...ed e g output of stable value In addition output by pressing print protocol button in main menu is also possible Information on the format of the data can be found in Appendix 18 1 If the Lims output...

Страница 31: ...n Units Instrument Instrument Name of Instrument MasterFw Firmware Version master board user Interface Chamber Name of measurement chamer ChamberSnr Serial number of measurement chamber equal to instr...

Страница 32: ...e chamber the entered batch name is reused 8 4 1 Configuration To activate the batch name input at measurement start go to Chamber set tings Data logger Batch name input The following options can be s...

Страница 33: ...d by No vasina If you have ordered the wireless barcode scanner together with a LabMaster aw neo Novasina took already care for the implementation and you don t have to proceed with the tasks listed b...

Страница 34: ...y board which can be found in the front module of the LabMaster aw neo as shown in the pictures below Once done mount the LabMaster aw neo front module back to the LabMaster aw neo housing 3 Connectin...

Страница 35: ...etc are logged and can neither be deleted nor changed by a user even if the user has full permissi on rights on the instrument The entire history of the instrument can be analyzed and also be exporte...

Страница 36: ...p Examples User Events U Log in log off Start of measurement Open measurement chamber Confirmed messages by user Parameter changes P Change measurement temperature Change password Back spacing filter...

Страница 37: ...ility Each pro duct type has its own stability settings In order to aviod time consuming repro gramming the LabMaster aw neo offers sample database capability In other words specific settings for each...

Страница 38: ...loading an existing method or creating a new one there is always a file operations window popping up The following options can be selected from in this window Save actual settings to sample database C...

Страница 39: ...permission profile The advantage of such permission profile user account set ups is obvious If a person quites the job there is no need to delete a complete permission set up Only the user account nee...

Страница 40: ...he permission profile which his account is allocated to A permission profile which is allocated to at least one user account cannot be deleted In this case a info is issued which user accounts are all...

Страница 41: ...new profile by hit New permission profile Delete an existing profile by pushing Delete permission profile Hints If a permission profile is modified modification are valid for all user accounts allocat...

Страница 42: ...ata to SD card PSED Permission to enter product name and sample num ber export data to SD card and delete log data All Full permission to edit and configure data logger Sensor calibration Permission t...

Страница 43: ...e Allocation of Permisssion Profile to User Account by selecting the appropriate permission profile Permission Profile of the cur rently logged in user cannot be changed If no permission profile has b...

Страница 44: ...not be recognized Thereby an optical or acoustic external transducer can be connected to the instrument and be configured to provide a recognisable signal in such environment The connector is a 230VA...

Страница 45: ...king on instrument with previous firmware versions The license code will not be overwritten once configuration is loaded 12 3 Reset to Factory Settings It can be necessary that a reset to factory set...

Страница 46: ...ill pop up and provide further details about a specific option or menu point 13 Maintenance The following chapter describes the main maintenance task which can be per formed by the user himself 13 1 C...

Страница 47: ...Such chemical protection filters have to be replaced periodically due to the fact that they get saturated As filters do not cover the IR sensor measurement time will not be elongated nor disturb the...

Страница 48: ...be replaced immediately 13 7 How to Change a Measurement Sensor If a measurement sensor is treated correctly by using appropriate chemical filter pro tection if necessary lifetime should be around 3...

Страница 49: ...Novasina AG Neuheimstrasse 12 8853 Lachen SWITZERLAND Tel 41 55 642 67 67 info novasina ch www novasina ch...

Отзывы: