5. How to Conduct a Measurement
5.1 Conduct a first Measurement
However, even if there is a word that the first steps are the most difficult and
challenging one, - honestly- forget about that. Doing a first measurement with
the LabMaster-aw neo is simple as 1-2-3. Just follow the step-by-step instruction
below:
1. Prepare sample in a way that the small portion is representative. Cut
the sample in small pieces, grind it, homogenise it or use another way
to create little chunks. Just do not use ceramics as it absorbs free
water and can change the water activity value
2. Fill sample cup half to 3/4. Do not overfill as it will just squeeze sample
into the protection filter which results in filter replacement or cleaning
work
3. Program important parameter as measurement temperature and
select appropriate stable mode / stable time. Feel unsure? Just read
chapter 4.
4. Go to main screen and press „START“ button
5. Measurement will be finished once stability has been reached. In that
moment, an acoustic indication (beep) is also released.
6. Open measurement chamber and remove sample. Always remove
sample once measurement is completed. NEVER switch off instrument
with sample or salt standards left in the chamber