4.1.2 Functionality of Temperature Control
• Sample surface temperature is the one which is regognized and control is
set to. Controller is optimizing itself to current operation and environmental
temperatures.
•
Sample surface temperature is kept constant to target value +/-0.02°C.
•
If instrument is operated in completely new conditions or target value is
changed remarkably, it can take up to 30mins until target value +/-0.30C is
reached. It is not an error of the instrument, it is more an indication that
optimization is in progress
•
Once optimization of control has been finished, the parameters are stored
once instrument is shut-down or temperature control is set to „OFF“