4.4. Commissioning
4. Preparing for use
SPECIFICATION OF THE REMOTE ON/OFF INTERFACE
• Labeled:
REMOTE
• Max. cable length: ?? m
1
2
3
4
Connector type of the Remote On/Off interface: Amphenol M8S-04PFFP-
SF8001 socket
Table 4.22.: Pin assignment of the Remote On/Off interface
Pin
Name
Description
Switch on: close pins 1 and 2
Switch off: open pins 1 and 2
1
B2B_SPS_REMOTE_ON
Power input 3.3 V
2
GND
Ground
While
on
pins 3 and 4 are internally
closed via an optocoupler
3
PWR_REMOTE_5V_0A1_08
Power output 5 V
4
B2B_SPS_REMOTE_RUN_N
(assumption; voltage may vary)
4.4. COMMISSIONING
4.4.1. CHANGING THE NETWORK SETTINGS
This procedure is necessary in order to connect the InScript software with the
ASC later on.
TARGET AUDIENCE AND QUALIFICATION
Experts
66
ASC_ARGNET-series_manual, 2022-02-15
Содержание ARGNET ASC-1