Note: The tip of the sensor may be protected by a yellow plastic cap to prevent damage in shipping. 
Remove the yellow cap before installing the sensor. 

The

 225-HMP60 

sensors must be protected from direct sunlight and precipitation when used outdoors 

for weather monitoring. A variety of sunshields are available, ranging from self-aspirated to fan-
aspirated models. Fan aspirated sunshields are preferred where there is little or no air movement. 
Some fan-aspirated units include a solar array, eliminating the need for an external power source for 
the fan.  

 

 

Note: When model 380-280-HMP60 Sunshield is ordered at the same time as the 255-
HMP60x Sensor, the sensor is pre-assembled, making installation simpler. 

 

Recommended sunshields include brackets for mast mounting by means of a U-bolt. The U-bolt can be 
removed for surface mounting the bracket. Refer to instructions included with the specific model for 
installation of the sensor and mounting options.  

Mast-mounted sensors are usually positioned at "shelter" height, or about four feet above ground 
level, on the north side of the mast. Temperature readings can be affected by heat reflected from the 
ground so try to avoid installation over asphalt or bare ground.  

Wire the sensor to your monitoring equipment according to the wire installation tables (Section 3). Use 
the table marked "225-HMP60 with Extended Cable" if the optional additional cable was ordered. 
Program your logger or PLC to power the sensor for at least 4 seconds before taking a reading. 

6

 

MAINTENANCE 

 

The

 225-HMP60 Relative Humidity and Temperature Sensors

 do not require calibration and have no 

calibration adjustments available to the user. Field-check the accuracy of the sensor by comparison 
with an accurate reference on an annual basis, or more often in dusty or polluted areas. 

The probe may be removed from the sunshield for inspection and service if necessary. With the probe 
removed, clean off any dust that has collected on or inside the sunshield. 

Sunshield

Type

Power Supply

380-280-HMP60

Self Aspirated

n/a

380-281

Self Aspirated

n/a

380-283

Fan Aspirated

Integrated solar panel / NiMh battery

380-RHRS

Self Aspirated

n/a

380-43502

Fan Aspirated

12-14 Vdc @ 500 mA for blower, AC adapter included

Содержание 225-HMP60A

Страница 1: ...225 HMP60x UM 20210202 225 HMP60x User Manual 225 HMP60A 0 1 V outputs 225 HMP60C 0 5 V outputs Temperature and Relative Humidity Sensor Phone 530 823 7185 Email nova novalynx com Website www novalynx...

Страница 2: ...e items If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packaging Please write the RA on the outside o...

Страница 3: ...Corporation 225 HMP60x Page 3 February 2021 CONTENTS 1 FORWARD 4 2 INTRODUCTION 4 3 WIRE INSTALLATION TABLE 4 4 SPECIFICATIONS 5 5 INSTALLATION 6 6 MAINTENANCE 6 7 CALIBRATION TEST KIT 7 8 TROUBLESHO...

Страница 4: ...or range is 0 to 100 relative humidity Temperature is measured by a 1000 ohm RTD element The temperature range is 40 to 60 C 40 to 140 F Model 225 HMP60A features two 0 1 V outputs Model 225 HMP60C fe...

Страница 5: ...HMP60A RL min 10 k HMP60C RL min 50 k Accuracy at 20 C 68 F 0 2 of FS Temperature dependence 0 01 of FS C 0 006 of FS F Power consumption 1 mA average max peak 5 mA Operating voltage HMP60A 5 28 VDC H...

Страница 6: ...e usually positioned at shelter height or about four feet above ground level on the north side of the mast Temperature readings can be affected by heat reflected from the ground so try to avoid instal...

Страница 7: ...or 7 CALIBRATION TEST KIT The NovaLynx 220 HMK15 Humidity Calibrator sold separately is available for checking the calibration of electronic RH sensors by the user The calibrator includes two chambers...

Отзывы: