The sensor elements are protected by a removable cap/filter. The cap can be replaced if the filter is 
dirty. The humidity sensor chip is replaceable. The chip is factory-trimmed so that there is no need to 
calibrate after replacing the part. The temperature sensing element is a separate part and is not user-
serviceable.

 Note: Unscrew counter-clockwise to remove the cap. 

 

 

 

The base of the probe is connected to the cable by an M8 connector. 

Note: The connection may be 

covered by heat shrink tubing.

 The probe may be disconnected and returned to NovaLynx for a 

calibration check, or the entire probe, cable and sunshield can be returned for service or a calibration 
check. If only the probe is to be returned, inspect the cable carefully to ensure it is not the cause of any 
problems you may have experienced with the sensor. 

7

 

CALIBRATION TEST KIT 

 

The NovaLynx 

220-HMK15 Humidity Calibrator

 (sold separately) is available for checking the 

calibration of electronic RH sensors by the user. The calibrator includes two chambers with saturated 
salt solutions to test the sensor at 12% and 75% RH. Testing at two points verifies the overall operating 
range of the humidity sensor. Spot checking at ambient conditions may not always detect a marginally 
defective sensing element. 

The calibration chambers work best in stable temperature conditions (indoors), but can also be used to 
make quick and accurate field tests. In either case, the chamber humidity reading is corrected 
according to the temperature indicated on a built-in thermometer. Sensor readings can be taken 
within 5 minutes.  

8

 

TROUBLESHOOTING 

 

If the sensor’s output signal appears to be in error or is absent, check the power connections. At the 
sensor cable, measure the input power source voltage with a voltmeter. Check each signal output by 
disconnecting the signal wire from the monitoring equipment and then measure the output voltage.

 

Содержание 225-HMP60A

Страница 1: ...225 HMP60x UM 20210202 225 HMP60x User Manual 225 HMP60A 0 1 V outputs 225 HMP60C 0 5 V outputs Temperature and Relative Humidity Sensor Phone 530 823 7185 Email nova novalynx com Website www novalynx...

Страница 2: ...e items If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packaging Please write the RA on the outside o...

Страница 3: ...Corporation 225 HMP60x Page 3 February 2021 CONTENTS 1 FORWARD 4 2 INTRODUCTION 4 3 WIRE INSTALLATION TABLE 4 4 SPECIFICATIONS 5 5 INSTALLATION 6 6 MAINTENANCE 6 7 CALIBRATION TEST KIT 7 8 TROUBLESHO...

Страница 4: ...or range is 0 to 100 relative humidity Temperature is measured by a 1000 ohm RTD element The temperature range is 40 to 60 C 40 to 140 F Model 225 HMP60A features two 0 1 V outputs Model 225 HMP60C fe...

Страница 5: ...HMP60A RL min 10 k HMP60C RL min 50 k Accuracy at 20 C 68 F 0 2 of FS Temperature dependence 0 01 of FS C 0 006 of FS F Power consumption 1 mA average max peak 5 mA Operating voltage HMP60A 5 28 VDC H...

Страница 6: ...e usually positioned at shelter height or about four feet above ground level on the north side of the mast Temperature readings can be affected by heat reflected from the ground so try to avoid instal...

Страница 7: ...or 7 CALIBRATION TEST KIT The NovaLynx 220 HMK15 Humidity Calibrator sold separately is available for checking the calibration of electronic RH sensors by the user The calibrator includes two chambers...

Отзывы: