Страница 3: ...Corporation 225 HMP60x Page 3 February 2021 CONTENTS 1 FORWARD 4 2 INTRODUCTION 4 3 WIRE INSTALLATION TABLE 4 4 SPECIFICATIONS 5 5 INSTALLATION 6 6 MAINTENANCE 6 7 CALIBRATION TEST KIT 7 8 TROUBLESHO...

Страница 4: ...or range is 0 to 100 relative humidity Temperature is measured by a 1000 ohm RTD element The temperature range is 40 to 60 C 40 to 140 F Model 225 HMP60A features two 0 1 V outputs Model 225 HMP60C fe...

Страница 5: ...HMP60A RL min 10 k HMP60C RL min 50 k Accuracy at 20 C 68 F 0 2 of FS Temperature dependence 0 01 of FS C 0 006 of FS F Power consumption 1 mA average max peak 5 mA Operating voltage HMP60A 5 28 VDC H...

Страница 6: ...e usually positioned at shelter height or about four feet above ground level on the north side of the mast Temperature readings can be affected by heat reflected from the ground so try to avoid instal...

Страница 7: ...or 7 CALIBRATION TEST KIT The NovaLynx 220 HMK15 Humidity Calibrator sold separately is available for checking the calibration of electronic RH sensors by the user The calibrator includes two chambers...