14. Practical Test 

 

An optimal correct pressure setting  permits the patient to sink into the mattress by 
about 20 to 30% of the original cell height. In alternating mode a hand or lower arm 
should be inserted effortlessly between patient and basic mattress into the respective 
ventilated cell region. Under no circumstances may the  patient hit the mattress 
bottom. 
 

 

15. System operation with one cell extracted: 

 

The LAS 500 mattress system permits continued operation with one cell extracted. 
Release the CPC coupling of the affected cell, remove the cell und lock the coupling 
with  a  CPC  sealing  plug  (optional,  item  no.:  991016).  Ensure  that  it  locks  in  place 
properly.  
Now the system can be used once again without difficulty. 
The extracted cell can in re-inserted in the opposite sequence. 
 
 

  

16. CPR - Valve : 

 

The

 novacare

 LAS 500 Low Air Loss Therapy System 

is equipped with a CPR 

emergency valve. It permits a spontaneous deflation for example to conduct 
reanimation. In this case set the valve to  switch position „open“. 
 
 
 

17. Caution note :  

 

• 

Protect control unit from humidity and direct wetness 

• 

Keep tubes free of bends 

• 

Do not use system in the immediate proximity of a source of heat. 

• 

When moving the bed pull out the power plug.  

• 

Never pull the cord.  

• 

Sharp  objects  must  be  kept  away.  Do  not  secure  overlay  with  inappropriate 

means. 

• 

Do  not  operate  system  in  the  immediate  proximity  of  flammable  gases  (risk  of 

explosion) 

• 

Use only original spare parts and consumables.  

• 

If  side  rails  are  used,  ensure  the  adherence  to  the  mandatory  minimum  

distance. Side rails need to be boosted if the  minimum distance cannot be met. 

 

 
 
 
 
 
 
 

Содержание LAS 500

Страница 1: ...Operating Instructions LAS 500 Low air Loss System...

Страница 2: ...tion Page 5 6 8 Lock Function Page 6 9 Backup Function Page 7 10 Static mode Page 7 11 Bariatric Heavy weight function Page 7 12 Care Function Page 7 13 Alarm function Page 7 14 Practical Test Page 8...

Страница 3: ...placement system for clinical stationary and ambulatory care taking The system consists of a control unit with digital operating unit and a mattress system with 18 arranged in a transverse manner 2 Se...

Страница 4: ...eight observe chart 2 Setting of alternating pressure interval Cycle 3 Alternating mode 4 Static mode 5 Care Mode 6 Start button Operate or Standby 7 display alternating pressure interval Cycle 5 95 i...

Страница 5: ...the bed using the fixation hooks The automatically retractable hooks adjust to the foot end of the bed and provide a secure hold Upon removal of the control unit the hooks automatically retract to the...

Страница 6: ...ments 0 95 Min by pushing the button Push the Alter Button to return to 5 12 Set patient weight by pushing the key Use setting chart for support 13 The patient can now be placed onto the mattress 14 N...

Страница 7: ...tric mattress optional in combination with this function The mode is activated by pushing the BARIATRIC Button The Bariatric LED lights up 12 Use of CARE Function The LAS 500 control unit features a C...

Страница 8: ...d once again without difficulty The extracted cell can in re inserted in the opposite sequence 16 CPR Valve The novacare LAS 500 Low Air Loss Therapy System is equipped with a CPR emergency valve It p...

Страница 9: ...with RL 93 42 EWG Use and applications are subject to these legal foundations Systems are suitable for reuse However hygiene directives must be observed and flawless operation must be ensured Mainten...

Страница 10: ...cribed by the respective manufacturer Guarantee claims become null and void if inappropriate and unlisted preparations are used or are used inappropriately We recommend our cleaning service Please ref...

Страница 11: ...al PU Max Flow 1200 ltr min Cycle Adjustable 5 95 Minutes in 5 Minute intervals Surrounding conditions Temperature Application operation 5 50 Transport storage 10 60 Rel humidity Application operation...

Страница 12: ...medical care products They are sold by novacare gmbh primarily to medical experts care taking personnel specialized medical dealers users medical facilities etc and put at the disposal of professional...

Страница 13: ...intervals prescribed by the manufacturer are adhered to As far as warranty claims are asserted and if during the subsequent inspection it is found that the damages are of wear and tear or are not subj...

Страница 14: ...LAS_500_eng page 14 31 Notes...

Страница 15: ...LAS_500_eng page 15...

Страница 16: ...acare de Copyright The instruction manual is protected by copyright Any reproduction and duplication recording in online services and the internet as well as any saving or copying on data carriers lik...

Отзывы: