LAS_500_eng
page 8
14. Practical Test
An optimal correct pressure setting permits the patient to sink into the mattress by
about 20 to 30% of the original cell height. In alternating mode a hand or lower arm
should be inserted effortlessly between patient and basic mattress into the respective
ventilated cell region. Under no circumstances may the patient hit the mattress
bottom.
15. System operation with one cell extracted:
The LAS 500 mattress system permits continued operation with one cell extracted.
Release the CPC coupling of the affected cell, remove the cell und lock the coupling
with a CPC sealing plug (optional, item no.: 991016). Ensure that it locks in place
properly.
Now the system can be used once again without difficulty.
The extracted cell can in re-inserted in the opposite sequence.
16. CPR - Valve :
The
novacare
LAS 500 Low Air Loss Therapy System
is equipped with a CPR
emergency valve. It permits a spontaneous deflation for example to conduct
reanimation. In this case set the valve to switch position „open“.
17. Caution note :
•
Protect control unit from humidity and direct wetness
•
Keep tubes free of bends
•
Do not use system in the immediate proximity of a source of heat.
•
When moving the bed pull out the power plug.
•
Never pull the cord.
•
Sharp objects must be kept away. Do not secure overlay with inappropriate
means.
•
Do not operate system in the immediate proximity of flammable gases (risk of
explosion)
•
Use only original spare parts and consumables.
•
If side rails are used, ensure the adherence to the mandatory minimum
distance. Side rails need to be boosted if the minimum distance cannot be met.