1.  The  System / Mode of action / area of applications  

 

 

The

 novacare

 LAS 500 Low Air Loss Therapy System 

is a true Low-air-Loss-

System“. It can be operated in static as well as in alternating mode.  
Its technology is based on a high air stream, in connection with large air openings 
of  the  cells.  The  result  is  an  air  cushion  effect.  It  enables  the  patient  to  sink  in 
deeply, resulting in a reduced pressure  Therefore the system is especially suited 
for the application on pain-sensitive patients.  
Contrary to ordinary alternating pressure system the alternating mode operates in 
a 3 to 1 principle. The system alternately ventilates two cells, while it removes air 
from  the  third  one.  This  results  in  a  very  low  partial  bearing  pressure  in  the 
alternate mode as well.  

The

  novacare

  LAS  500  Low  Air  Loss  Therapy  System   

is  a  mattress 

replacement system for clinical, stationary and ambulatory care-taking.  
The  system  consists  of  a  control  unit  with  digital  operating  unit  and  a  mattress 
system with 18 arranged in a transverse manner.  
 
 

2. 

Severity of pressure ulcers

 

The 

novacare

  LAS  500  Low  Air  Loss  Therapy  System 

is  recommended  for 

pressure ulcer treatment up to decubitus degree 4 (EPUAP)  / high risk to very high 
risk.  
 
 

3.  Maximum recommended patient weight  

 

 

In  standard  operation  the 

novacare

  LAS  500  Low  Air  Loss  Therapy 

System 

is recommended for a maximum patient weight of 

180 kg

.  

 

An  additional  bariatric-  /  increased  function  permits  applications  for  a  patient 
weight up to 

320 kg.

 For extra comfort we recommend the use of an optional 

120cm wide bariatric mattress. 

 

 

4.  Indications  / Contraindications : 

Indications 

Treatment  and  prevention  of  pressure  ulcers  and  decubitus 
especially for patients at high risk, as well as for heavy weight 
patients. 

 

Contraindications: 

Alternating  pressure  should  be  applied  to  pain  or  pain-
sensitive patients or in case of instable spinal cord injuries! 

 

 

 

 
 
 

Содержание LAS 500

Страница 1: ...Operating Instructions LAS 500 Low air Loss System...

Страница 2: ...tion Page 5 6 8 Lock Function Page 6 9 Backup Function Page 7 10 Static mode Page 7 11 Bariatric Heavy weight function Page 7 12 Care Function Page 7 13 Alarm function Page 7 14 Practical Test Page 8...

Страница 3: ...placement system for clinical stationary and ambulatory care taking The system consists of a control unit with digital operating unit and a mattress system with 18 arranged in a transverse manner 2 Se...

Страница 4: ...eight observe chart 2 Setting of alternating pressure interval Cycle 3 Alternating mode 4 Static mode 5 Care Mode 6 Start button Operate or Standby 7 display alternating pressure interval Cycle 5 95 i...

Страница 5: ...the bed using the fixation hooks The automatically retractable hooks adjust to the foot end of the bed and provide a secure hold Upon removal of the control unit the hooks automatically retract to the...

Страница 6: ...ments 0 95 Min by pushing the button Push the Alter Button to return to 5 12 Set patient weight by pushing the key Use setting chart for support 13 The patient can now be placed onto the mattress 14 N...

Страница 7: ...tric mattress optional in combination with this function The mode is activated by pushing the BARIATRIC Button The Bariatric LED lights up 12 Use of CARE Function The LAS 500 control unit features a C...

Страница 8: ...d once again without difficulty The extracted cell can in re inserted in the opposite sequence 16 CPR Valve The novacare LAS 500 Low Air Loss Therapy System is equipped with a CPR emergency valve It p...

Страница 9: ...with RL 93 42 EWG Use and applications are subject to these legal foundations Systems are suitable for reuse However hygiene directives must be observed and flawless operation must be ensured Mainten...

Страница 10: ...cribed by the respective manufacturer Guarantee claims become null and void if inappropriate and unlisted preparations are used or are used inappropriately We recommend our cleaning service Please ref...

Страница 11: ...al PU Max Flow 1200 ltr min Cycle Adjustable 5 95 Minutes in 5 Minute intervals Surrounding conditions Temperature Application operation 5 50 Transport storage 10 60 Rel humidity Application operation...

Страница 12: ...medical care products They are sold by novacare gmbh primarily to medical experts care taking personnel specialized medical dealers users medical facilities etc and put at the disposal of professional...

Страница 13: ...intervals prescribed by the manufacturer are adhered to As far as warranty claims are asserted and if during the subsequent inspection it is found that the damages are of wear and tear or are not subj...

Страница 14: ...LAS_500_eng page 14 31 Notes...

Страница 15: ...LAS_500_eng page 15...

Страница 16: ...acare de Copyright The instruction manual is protected by copyright Any reproduction and duplication recording in online services and the internet as well as any saving or copying on data carriers lik...

Отзывы: