9. Backup – Function :
In case of power loss, the settings used last are saved. New adjustments after
power loss are not necessary.
10. Static mode:
Set the mode button to „Static” (LED – display „STATIC“ lights up).
The unit can be operated in static mode for special applications (i.e. pain
patients)
11. Bariatric- / Heavy weight function :
The
novacare
LAS 500 Heavy weight therapy systems
features a „Bariatric-
Function accommodating especially large patients. We recommend the use of
the bariatric mattress (optional) in combination with this function.
The mode is activated by pushing the BARIATRIC – Button. The “Bariatric” LED
lights up.
12. Use of CARE Function:
The LAS 500 control unit features a CARE Function. By
activating this function all chambers of the mattress system are
inflated with maximum system pressure for 30 min. After 30
min., for the safety of the patient, the system automatically
switches to the last selected mode.
This special function facilitates re-positioning of the patient during
care-taking.
13. Alarm function :
The unit features an acoustic power loss alarm. Power loss is signaled by intermittent
beeping.
This can be eliminated by turning the unit off (ON/OFF – Switch).