54
Cleaning and Disinfecting the Meter
NOTE:
To properly clean and disinfect the Meter, steps 1 to 5 should
be performed together before testing on each patient.
1. Clean the Meter.
• Wipe the external surface of the Meter thoroughly with a fresh
Germicidal Wipe.
• Discard the used wipe per Step 4.
2. Disinfect the Meter.
• Remove another fresh Germicidal Wipe from the
canister. Thoroughly wipe the top, bottom, left,
and right sides of the Meter avoiding the Test
Strip port by wiping the surface a minimum of 3
times horizontally followed by 3 times vertically.