33
NOTE:
If the strip is removed before the test starts or is not used for
over 2 minutes, the screen will go blank.
NOTE:
A quick beep sounds when sufficient Control Solution has
been added to the Test Strip.
NOTE:
Nova Max Pro Creat eGFR Control Solution test results are
available on-screen in 30 seconds.
Out-of-range results may be caused by the following:
• You may not be doing the test properly. Retest and follow the
instructions carefully.
• The control solution may have expired or have been con taminated.
Check the expiration date on the control solution vial. The control
solution is good for only 90 days after opening. Make sure the
control solution vial is closed when not in use.
Running Nova Max Pro Creat eGFR Control Solution
eGFR MeterIFU Manual PRO PN64322.indd 33
eGFR MeterIFU Manual PRO PN64322.indd 33