Perform a Control Solution Test
The Control Solution test confirms that the Meter and Test Strips
are working correctly. A Control Solution test is similar to a blood
test, except the Nova Max Pro Creat eGFR Control Solution is used
instead of a blood sample.
When to Perform a Quality Control Test
• Before using the Nova Max Pro Creat and eGFR Meter for the first
time.
• As required by your institution's quality control policy or local
regulatory requirements.
• If a patient test has been repeated and the blood Creatinine results
are still lower or higher than expected.
• If there are other indications that the system is not working properly.
• If you drop the meter.
Running Nova Max Pro Creat eGFR Control Solution