• The test strip may have expired. Check the expiration date on the
test strip vial.
• The test strip may have been damaged. Retest using a new test
strip.
• The Nova Max Pro Creat eGFR Meter may not be working properly.
• Acceptable quality control ranges may differ at your institu tion
from the ranges reported by Nova Biomedical. Follow your institu-
tion’s guidelines for quality control results.
NOTE:
If the Control Solution test result is outside the range (is either
higher or lower), the Meter and Test Strip may not be working as a
system. Repeat the process using a new Test Strip. Do not use the Meter
until test results fall within the appropriate range. If the problem contin-
ues, call Customer Service.
Nova Max Pro Creat eGFR Control/Linearity Material are traceable to
NIST Standard Reference Material.