NC iQ
30 EN
After the expiry of the product life cycle:
E
Dispose of the product, displaying an icon as shown to the side, according to the
applicable regulations for electronic waste. Check with your authorised dealer or
with the responsible authorities about the applicable regulations.
E
Provide a container at the workstation for the proper disposal of contaminated
parts.
E
Discard used or defective needle modules, as well as needle modules in damaged packag-
ing in a puncture-proof container (Sharps container) according to the current regulations
of your country.
10 IN THE EVENT OF QUESTIONS AND PROBLEMS
The following malfunctions or error messages may occur:
FAULT/
MESSAGE
FAULT
CAUSE
ACTION TAKEN
“Handpiece
detached!“
The hand-
piece is not
detected.
The connection cable is
not properly connected
E
Carry out all the work steps in this
chapter 5.4 "
Connecting the hand-
piece
“ on page 20.
“System
Overloaded
Press any
Key!“
There is a
fault signal at
the hand-
piece.
The connection cable is
not properly connected.
E
Check the connections between the
handpiece and the control unit.
E
Press any button on the control unit to
return to the default screen.
The needle module is
not properly installed.
E
Check the fit of the needle module in
the handpiece.
E
Press any button on the control unit to
return to the default screen.
“System Er-
ror - Restart
Device”
An error has
occurred at
the handpiece
or the control
unit.
E
Switch the control unit to the standby
mode and wait 30 seconds (see chap-
ter 6.1 on page 20).
E
Switch the control unit back into oper-
ating mode.
The control
unit does
not switch
into standby
mode.
The power
supply has
been discon-
nected.
The power supply
plug is not properly
connected.
E
Check the connection of the power
supply plug to the control unit (see
chapter 5.2 "
Connecting the hand-
piece
“ on page 18).
E
Check the mains power supply to the
workstation.