NC iQ
To turn the handpiece off:
E
Press the
RUN
button again ( ).
7
CLEANING AND MAINTENANCE
CAUTION
Short circuiting hazard and risk of slight electric shock does exist.
If components of the device are dismantled and cleaned while the unit is turned on, you run
the risk of incurring a slight electrical shock or damaging the electronics in the drive motor.
E
Disconnect the control unit from the power supply before beginning any cleaning and
maintenance work.
ATTENTION
Damage resulting from moisture/liquids
If cleaning or disinfecting agents should enter the inside of the handpiece, the control unit
or the power supply plug, it can cause short circuiting. In addition, the mechanical and elec-
trical components may corrode.
E
Never soak the handpiece, control unit or the power supply plug in cleaning or disin-
fectant agents.
E
Never clean the handpiece, the control unit or the power supply plug in an autoclave or
an ultrasonic bath.
E
When selecting cleaning and disinfecting agents, note material compatibilities (see
chapter 7.2 "
Material compatibilities
“ on page 28).
7.1 Inspection
An inspection may only be carried out by an authorised dealer.
We recommend to drop off the device at an authorised service dealer at regular 24 month
intervals for an inspection according to the manufacturer's specifications.