10.2.4
Although the EAS is a NIOSH approved accessory, NIOSH does not approve the use of an
EAS device. Use of the EAS voids NIOSH approval while being used.
10.3
FIRST BREATH CDV
10.3.1
The CDV with First Breath allows the user to save air. Open the cylinder valve and don the
facepiece. Activate the CDV by inhaling (taking a normal breath) and the button will release,
starting the flow of air. Prior to removing the SCBA, simply press the button inward and the
airflow will stop (Figure 42).
11.0 TROUBLE SHOOTING GUIDE
12.0 MAJOR COMPONENTS
12.1
Full Facepiece Assemblies
North Catalog Number
Description
903010
Complete Facepiece Assembly, Medium/Large
903010S
Complete Facepiece Assembly, Small
904010
Complete AirHood Assembly, Polyurethane, Large/ X Large
904010S
Complete AirHood Assembly, Polyurethane, Small/Medium
905010
Complete AirHood Assembly, PVC, Large/ X Large
905010S
Complete AirHood Assembly, PVC, Small/ Medium
12.2
Cylinders
North Catalog Number
Description
910301LP
30 minute, Replacement Cylinder, Aluminum, 2216 psi
910302LP
30 minute, Replacement Cylinder, Hoop Wrapped
Aluminum, 2216 psi
910303LP
30 minute, Replacement Cylinder, Full Wrapped Carbon,
2216 psi
910302HP
30 minute, Replacement Cylinder, Hoop Wrapped
Aluminum, 4500 psi
22
NOTE
If using the AirHood with a First Breath CDV, after donning immediately push the seal of the
oral/nasal cup to the face and inhale in order to activate the first breath. A quick rush of air will be
heard when airflow is activated.
No air flow inside the facepiece
1.
Cylinder valve is not open
1.
Open the cylinder valve
2.
No air in the cylinder
2.
Fill the cylinder
Low air flow inside the facepiece
Cylinder valve is only slightly open
Open cylinder valve several turns
Air free flows in facepiece at all times
The Bypass is open
Close the Bypass
Air leakage is heard when exhaling
The Bypass is slightly open
Close the Bypass fully
PROBLEM
POSSIBLE CAUSE
CORRECTION