User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
57 -
11.2 Other Terms and Messages
“Cuff error”
-
cuff is not wrapped correctly, or is not connected
“Air leak”
-
Pneumatic component, tubing or cuff leaks
“Pressure error”
-
Unstable cuff pressure or tangled cuff tubing
“Signal weak”
-
Very weak signal because of the cuff, or the patient has very weak pulse
“Over extent”
-
The measurement range exceeds 255 mmHg (Infant/Neonatal 135 mmHg)
“Over motion”
-
The repeated measurement due to moving, excessive noise during the stepping inflation
andmeasuring pressure and pulse, e.g. during patient shaking motion
“Signal overflow”
-
Blood pressure amplifier overflow due to excessive movement
“Leak in gas run”
-
Leaking during the pneumatic device testing
“System error”
-
Abnormal condition of CPU, such as register overflow, divided by zero
“Adult”
-
The blood pressure measuring now is in adult mode. In this case, it is not allowed to monitoring Infant
or Neonatal patient. Otherwise, there may be serious danger to the patient monitored.
“Infant”
-
The blood pressure module is now worked in infant measuring mode.
“LEADS OFF”
-
SpO
2
probe fell off
“PROBE OFF”
-
The ECG electrodes or cable fell off
“LEARNING”
-
Learning arrhythmia for 15 seconds
“DEMO”
-
The instrument is working at demonstration mode, all signals are produced inside the instrument.