User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
43 -
5.
Overpressure protection limit
Adult: 300 mmHg
pediatric: 300mmHg
Neonate: 140 mmHg
6.
NIBP measurement range:
Systolic
:
20 mmHg
〜
290 mmHg
Diastolic
:
10 mmHg
〜
260 mmHg
MAP
:
15 mmHg
〜
275 mmHg
7.
NIBP accuracy:
Systolic
:
± 5mmHg or ±3%
Diastolic
:
±5 mmHg or ±3%
MAP
:
±5 mmHg or ±3%
8.
Measurement mode: Manual, Auto
9.
Measurement Method: Adult, Pediatric, Neonate
10.
Automatic Cycling Test: 1~240min, Tolerance:<10s
6.5
SpO
2
Monitoring
1.
Transducer: Dual-wavelength LED
2.
SpO
2
measuring range: 0%~100%
3.
SpO
2
measuring accuracy: Arms is not greater than 2% for SpO
2
range from 70% to 100%
4.
SpO
2
measurement accuracy
:
(1) 70%
〜
100% ±2%
(2) Below 70%, the value is only for reference without exact definition, the symbol for % means the
“SpO
2
Percentage”.
5.
PI range: 0.4% ~ 20%
6.6
Pulse Rate monitoring
1.
Pulse rate measuring range: 10bpm~300bpm
2.
Pulse rate measuring accuracy: ±2bpm or ±2%, whichever is greater.
6.7
Data Recording
1.
Sensitivity selection tolerance:
≤5%
2.
Recording speed: 25mm/s
3.
Recording speed accuracy:
≤5%
6.8
Other Technical Specifications
1.
Power supply: 100~240VAC, 50/60Hz
2.
Power consumption: <100VA
3.
Display mode: 10.4 inches TFT color LCD
4.
Alarming mode: Audible & visible alarm
5.
Communication: NET(RJ45)