User Manual for Multi-parameter Patient Monitor North-vision Tech. Inc.
-
31 -
4.9.5 Temperature Parameter Settings
TEMP Hi :
Temperature 1 upper alarm setting, adjustable 0.1
– 60.0°C (32.1 - 140.0°F)
Lo :
Temperature 1 lower alarm setting, adjustable 0
– 59.9°C (32.0 - 139.8°F)
Unit
:
Temperature unit. The factory setting is °C (Celsius) and can be set to °F (Fahrenheit).
4.9.6 NIBP Parameter Settings
SYS Hi/Lo:
Systolic pressure upper and lower alarm limits.
DIA Hi/Lo:
Diastolic pressure upper and lower alarm limits.
MAP Hi/Lo:
Mean arterial pressure upper and lower alarm limits.
PR Hi/Lo:
Pulse rate upper and lower alarm limits.
Unit:
the pressure unit. mmHg or kPa. The factory default is mmHg.
Mode:
The cuff inflation mode, manual or automatic. The factory default is manual. The operator
needs to press the NIBP button to perform NIBP measurement.
>>
:
Next page.
Nibp Cail.:
This option is
for the provision of
factory
calibration
of
the
machine
,
"OFF
",
"
Mode
1"
and
"
Mode
2
".
The
factory
default value: "
Off".
Figure 4.33 NIBP Settings
Figure 4.32 Temperature settings